Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument
This research in determining x-ray effective energy in linear accelerator (Linac) radiotherapy device was done using copper plates at different thicknesses by determining the Half Value Layer (HVL) for each energy. The purpose of this research is to know the amount of effective energy used in radiot...
Main Authors: | , |
---|---|
Format: | Article |
Language: | Indonesian |
Published: |
Institut Agama Islam Negeri Raden Intan Lampung
2018-04-01
|
Series: | Jurnal Ilmiah Pendidikan Fisika Al-Biruni |
Subjects: | |
Online Access: | http://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/2336 |
Summary: | This research in determining x-ray effective energy in linear accelerator (Linac) radiotherapy device was done using copper plates at different thicknesses by determining the Half Value Layer (HVL) for each energy. The purpose of this research is to know the amount of effective energy used in radiotherapy linear accelerator. The results showed that the energy of 6 MeV obtained 2,714 MeV of effective energy at HVL 4,500 mm and at the energy of 10 MeV obtained 4,231 MeV of effective energy at HVL 4,500 mm. |
---|---|