Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument

This research in determining x-ray effective energy in linear accelerator (Linac) radiotherapy device was done using copper plates at different thicknesses by determining the Half Value Layer (HVL) for each energy. The purpose of this research is to know the amount of effective energy used in radiot...

Full description

Bibliographic Details
Main Authors: Fauziah Fauziah, Bualkar Abdullah
Format: Article
Language:Indonesian
Published: Institut Agama Islam Negeri Raden Intan Lampung 2018-04-01
Series:Jurnal Ilmiah Pendidikan Fisika Al-Biruni
Subjects:
Online Access:http://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/2336
Description
Summary:This research in determining x-ray effective energy in linear accelerator (Linac) radiotherapy device was done using copper plates at different thicknesses by determining the Half Value Layer (HVL) for each energy. The purpose of this research is to know the amount of effective energy used in radiotherapy linear accelerator. The results showed that the energy of 6 MeV obtained 2,714 MeV of effective energy at HVL 4,500 mm and at the energy of 10 MeV obtained 4,231 MeV of effective energy at HVL 4,500 mm.