Design of a Selective and Sensitive PVC-Membrane Potentiometric Sensor for Strontium Ion Based on 1,10-Diaza-5,6-benzo-4,7-dioxacyclohexadecane-2,9-dioneas a Neutral Ionophore
A novel PVC membrane sensor for the Sr2+ ion based on 1,10-diaza-5,6-benzo-4,7- dioxacyclohexadecane-2,9-dione has been prepared. The sensor possesses a Nernstian slope of 30.0 ± 0.6 mV decade-1 over a wide linear concentration range of 1.6 × 10-6-3.0 ×10-3 M with a detection limit of 6.3 Ã...
Main Authors: | Hashem Sharghi, Sayed Yahya Kazemi, Mojtaba Shamsipur |
---|---|
Format: | Article |
Language: | English |
Published: |
MDPI AG
2007-04-01
|
Series: | Sensors |
Subjects: | |
Online Access: | http://www.mdpi.com/1424-8220/7/4/438/ |
Similar Items
-
Ionophore-based potentiometric PVC membrane sensors for determination of phenobarbitone in pharmaceutical formulations
by: Alrabiah Haitham, et al.
Published: (2016-12-01) -
PVC Membrane Sensors for Potentiometric Determination of Acebutolol
by: Abdulrahman Al-Majed, et al.
Published: (2007-12-01) -
Efficient Synthesis of a Range of Benzosubstituted Macrocyclic Diamides
by: Hashem Sharghi, et al.
Published: (2000-06-01) -
Acyclic Arylamine-Based Ionophores as Potentiometric Sensors for Zn2+ and Ni2+ Ions
by: Harpreet Kaur, et al.
Published: (2017-11-01) -
PVC-membrane potentiometric sensors based on a recently synthesized Schiff base for Fe(III) ion
by: S. Yousef Ebrahimipur, et al.
Published: (2012-12-01)