Decontamination of digital image sensors and assessment of electron microscope performance in a BSL-3 containment

A unique biological safety level (BSL)-3 cryo-electron microscopy facility with a 200 keV high-end cryo-electron microscope has been commissioned at the University of Texas Medical Branch (UTMB) to study the structure of viruses and bacteria classified as select agents. We developed a microscope dec...

Full description

Bibliographic Details
Main Authors: Michael B. Sherman, Juan Trujillo, Benjamin E. Bammes, Liang Jin, Matthias W. Stumpf, Scott C. Weaver
Format: Article
Language:English
Published: AIMS Press 2015-05-01
Series:AIMS Biophysics
Subjects:
Online Access:http://www.aimspress.com/biophysics/article/285/fulltext.html
id doaj-8f5e0bf2888e403caad528ccee656a17
record_format Article
spelling doaj-8f5e0bf2888e403caad528ccee656a172020-11-25T00:19:04ZengAIMS PressAIMS Biophysics2377-90982015-05-012215316210.3934/biophy.2015.2.153201502153Decontamination of digital image sensors and assessment of electron microscope performance in a BSL-3 containmentMichael B. ShermanJuan Trujillo0Benjamin E. Bammes1Liang Jin2Matthias W. Stumpf3Scott C. WeaverJEOL USA, Inc., Peabody, Massachusetts, USADirect Electron, LP, San Diego, California, USADirect Electron, LP, San Diego, California, USAE. A. Fischione Instruments, Inc., Export, Pennsylvania, USAA unique biological safety level (BSL)-3 cryo-electron microscopy facility with a 200 keV high-end cryo-electron microscope has been commissioned at the University of Texas Medical Branch (UTMB) to study the structure of viruses and bacteria classified as select agents. We developed a microscope decontamination protocol based on chlorine dioxide gas with a continuous flow system. In this paper we report on testing digital camera sensors (both CCD and CMOS direct detector) in a BSL-3 environment, and microscope performance after chlorine dioxide (ClO<sub>2</sub>) decontamination cycles.http://www.aimspress.com/biophysics/article/285/fulltext.htmlcryo-electron microscopysingle particle imagingbiological safety containment
collection DOAJ
language English
format Article
sources DOAJ
author Michael B. Sherman
Juan Trujillo
Benjamin E. Bammes
Liang Jin
Matthias W. Stumpf
Scott C. Weaver
spellingShingle Michael B. Sherman
Juan Trujillo
Benjamin E. Bammes
Liang Jin
Matthias W. Stumpf
Scott C. Weaver
Decontamination of digital image sensors and assessment of electron microscope performance in a BSL-3 containment
AIMS Biophysics
cryo-electron microscopy
single particle imaging
biological safety containment
author_facet Michael B. Sherman
Juan Trujillo
Benjamin E. Bammes
Liang Jin
Matthias W. Stumpf
Scott C. Weaver
author_sort Michael B. Sherman
title Decontamination of digital image sensors and assessment of electron microscope performance in a BSL-3 containment
title_short Decontamination of digital image sensors and assessment of electron microscope performance in a BSL-3 containment
title_full Decontamination of digital image sensors and assessment of electron microscope performance in a BSL-3 containment
title_fullStr Decontamination of digital image sensors and assessment of electron microscope performance in a BSL-3 containment
title_full_unstemmed Decontamination of digital image sensors and assessment of electron microscope performance in a BSL-3 containment
title_sort decontamination of digital image sensors and assessment of electron microscope performance in a bsl-3 containment
publisher AIMS Press
series AIMS Biophysics
issn 2377-9098
publishDate 2015-05-01
description A unique biological safety level (BSL)-3 cryo-electron microscopy facility with a 200 keV high-end cryo-electron microscope has been commissioned at the University of Texas Medical Branch (UTMB) to study the structure of viruses and bacteria classified as select agents. We developed a microscope decontamination protocol based on chlorine dioxide gas with a continuous flow system. In this paper we report on testing digital camera sensors (both CCD and CMOS direct detector) in a BSL-3 environment, and microscope performance after chlorine dioxide (ClO<sub>2</sub>) decontamination cycles.
topic cryo-electron microscopy
single particle imaging
biological safety containment
url http://www.aimspress.com/biophysics/article/285/fulltext.html
work_keys_str_mv AT michaelbsherman decontaminationofdigitalimagesensorsandassessmentofelectronmicroscopeperformanceinabsl3containment
AT juantrujillo decontaminationofdigitalimagesensorsandassessmentofelectronmicroscopeperformanceinabsl3containment
AT benjaminebammes decontaminationofdigitalimagesensorsandassessmentofelectronmicroscopeperformanceinabsl3containment
AT liangjin decontaminationofdigitalimagesensorsandassessmentofelectronmicroscopeperformanceinabsl3containment
AT matthiaswstumpf decontaminationofdigitalimagesensorsandassessmentofelectronmicroscopeperformanceinabsl3containment
AT scottcweaver decontaminationofdigitalimagesensorsandassessmentofelectronmicroscopeperformanceinabsl3containment
_version_ 1725373428659126272