%0 Article %A Clifton Carey %I MDPI AG %D 2015 %G English %B Chemosensors %@ 2227-9040 %T Plasticizer Effects in the PVC Membrane of the Dibasic Phosphate Selective Electrode %U http://www.mdpi.com/2227-9040/3/4/284 %X The PVC membrane of an ion-selective electrode (ISE) sensitive to dibasic phosphate ions (HPO4-ISE) has not been optimized for maximum selectivity, sensitivity, and useable ISE lifetime and further work was necessary to improve its performance. Two areas of investigation are reported here: include the parameters for the lipophilicity of the plasticizer compound used and the amount of cyclic polyamine ionophore incorporated in the PVC membrane. Six candidate plasticizers with a range of lipophilicity were evaluated for their effect on the useable lifetime, sensitivity, and selectivity of the ISE against 13 different anions. Selectivity was determined by a modified fixed interferent method, sensitivity was determined without interferents, and the usable lifetime evaluated at the elapsed time where 50% of the HPO4-ISE failed (L50). The results show that choosing a plasticizer that has a lipophilicity similar to the ionophore’s results in the best selectivity and sensitivity and the longest L50.