The shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick test

The global application of the skin prick test (SPT) is attributed to the low costs, easy execution, and in vivo approach. Still, the healthcare professionals’ technique and the lancet shape may challenge the standardization of the method. Thus, we investigated the influence of the shape of the lance...

وصف كامل

التفاصيل البيبلوغرافية
الحاوية / القاعدة:World Allergy Organization Journal
المؤلفون الرئيسيون: Ewa Anna Bartko, PhD, Jesper Elberling, PhD, MD, Holger Mosbech, DMSc, MD
التنسيق: مقال
اللغة:الإنجليزية
منشور في: Elsevier 2024-06-01
الموضوعات:
الوصول للمادة أونلاين:http://www.sciencedirect.com/science/article/pii/S1939455124000462
_version_ 1850287288579260416
author Ewa Anna Bartko, PhD
Jesper Elberling, PhD, MD
Holger Mosbech, DMSc, MD
author_facet Ewa Anna Bartko, PhD
Jesper Elberling, PhD, MD
Holger Mosbech, DMSc, MD
author_sort Ewa Anna Bartko, PhD
collection DOAJ
container_title World Allergy Organization Journal
description The global application of the skin prick test (SPT) is attributed to the low costs, easy execution, and in vivo approach. Still, the healthcare professionals’ technique and the lancet shape may challenge the standardization of the method. Thus, we investigated the influence of the shape of the lancet and the applied weight on the wheal size of SPT. Two allergic and one non-allergic individual were tested with allergens (Dermatophagoides pteronyssinus and Phleum pratense) and histamine solution (positive control), respectively. Horizontally (HS) and diagonally (DS) shouldered lancets with the same tip length (1 mm) were tested under two different conditions: either 60 g or 120 g weight pressure. The wheal size induced by the 4 different combinations was measured. The higher-weight device (120 g) induced a significantly larger and less variable wheal response with the tested allergens and histamine. However, the shape of the lancet affected the wheal size more than the applied weight. The least variable response was measured to histamine for the horizontal-shouldered lancet combined with the higher weight, whereas the same lancet with the lower weight resulted in a significant number of false negative results.
format Article
id doaj-art-6efca4b6a9e648b48d8befbe113b56ea
institution Directory of Open Access Journals
issn 1939-4551
language English
publishDate 2024-06-01
publisher Elsevier
record_format Article
spelling doaj-art-6efca4b6a9e648b48d8befbe113b56ea2025-08-19T23:36:58ZengElsevierWorld Allergy Organization Journal1939-45512024-06-0117610091510.1016/j.waojou.2024.100915The shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick testEwa Anna Bartko, PhD0Jesper Elberling, PhD, MD1Holger Mosbech, DMSc, MD2Corresponding author. Gentofte Hospital, Skin and Allergy Department, Allergy Clinic, Gentofte Hospitalsvej 22, 2900 Hellerup, Denmark.; Gentofte Hospital, Skin and Allergy Department, Allergy Clinic, Hellerup, DenmarkGentofte Hospital, Skin and Allergy Department, Allergy Clinic, Hellerup, DenmarkGentofte Hospital, Skin and Allergy Department, Allergy Clinic, Hellerup, DenmarkThe global application of the skin prick test (SPT) is attributed to the low costs, easy execution, and in vivo approach. Still, the healthcare professionals’ technique and the lancet shape may challenge the standardization of the method. Thus, we investigated the influence of the shape of the lancet and the applied weight on the wheal size of SPT. Two allergic and one non-allergic individual were tested with allergens (Dermatophagoides pteronyssinus and Phleum pratense) and histamine solution (positive control), respectively. Horizontally (HS) and diagonally (DS) shouldered lancets with the same tip length (1 mm) were tested under two different conditions: either 60 g or 120 g weight pressure. The wheal size induced by the 4 different combinations was measured. The higher-weight device (120 g) induced a significantly larger and less variable wheal response with the tested allergens and histamine. However, the shape of the lancet affected the wheal size more than the applied weight. The least variable response was measured to histamine for the horizontal-shouldered lancet combined with the higher weight, whereas the same lancet with the lower weight resulted in a significant number of false negative results.http://www.sciencedirect.com/science/article/pii/S1939455124000462Skin prick testAllergenPrick test lancetStandardizationMethodology
spellingShingle Ewa Anna Bartko, PhD
Jesper Elberling, PhD, MD
Holger Mosbech, DMSc, MD
The shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick test
Skin prick test
Allergen
Prick test lancet
Standardization
Methodology
title The shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick test
title_full The shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick test
title_fullStr The shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick test
title_full_unstemmed The shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick test
title_short The shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick test
title_sort shape of and applied weight on skin prick lancets critically affect the wheal size in the skin prick test
topic Skin prick test
Allergen
Prick test lancet
Standardization
Methodology
url http://www.sciencedirect.com/science/article/pii/S1939455124000462
work_keys_str_mv AT ewaannabartkophd theshapeofandappliedweightonskinpricklancetscriticallyaffectthewhealsizeintheskinpricktest
AT jesperelberlingphdmd theshapeofandappliedweightonskinpricklancetscriticallyaffectthewhealsizeintheskinpricktest
AT holgermosbechdmscmd theshapeofandappliedweightonskinpricklancetscriticallyaffectthewhealsizeintheskinpricktest
AT ewaannabartkophd shapeofandappliedweightonskinpricklancetscriticallyaffectthewhealsizeintheskinpricktest
AT jesperelberlingphdmd shapeofandappliedweightonskinpricklancetscriticallyaffectthewhealsizeintheskinpricktest
AT holgermosbechdmscmd shapeofandappliedweightonskinpricklancetscriticallyaffectthewhealsizeintheskinpricktest