An Experimental Setup for Power Loss Measurement up to 1 kHz using an Epstein Frame at CMI
This paper describes an experimental setup used at the Czech Metrology Institute (CMI) to measure the specific power loss of oriented and non-oriented electrical steel sheets up to 1 kHz using an Epstein frame. Special attention is given to a) a description of the hardware that is used, b) a descrip...
| Published in: | Measurement Science Review |
|---|---|
| Main Author: | |
| Format: | Article |
| Language: | English |
| Published: |
Sciendo
2023-12-01
|
| Subjects: | |
| Online Access: | https://doi.org/10.2478/msr-2023-0035 |
| Summary: | This paper describes an experimental setup used at the Czech Metrology Institute (CMI) to measure the specific power loss of oriented and non-oriented electrical steel sheets up to 1 kHz using an Epstein frame. Special attention is given to a) a description of the hardware that is used, b) a description of the feedback control and measurement software, and c) an analysis of the sources of uncertainty and validation. Calibration expanded uncertainty of (0.5 up to 1.6)% for k = 2 can be achieved with this setup. |
|---|---|
| ISSN: | 1335-8871 |