Voltammetric Determination of Carcinogenic Nitrobiphenyls at a Hanging Mercury Drop Electrode
Differential pulse voltammetry (DPV) and adsorptive stripping voltammetry (AdSV) at a hanging mercury drop electrode (HMDE) was used for the determination of trace amounts of carcinogenic nitrobiphenyls, namely 2-nitrobiphenyl (2-NBP), 3-nitrobiphenyl (3-NBP) and 4-nitrobiphenyl (4-NBP) within the c...
| الحاوية / القاعدة: | Sensors |
|---|---|
| المؤلفون الرئيسيون: | , , , |
| التنسيق: | مقال |
| اللغة: | الإنجليزية |
| منشور في: |
MDPI AG
2003-02-01
|
| الموضوعات: | |
| الوصول للمادة أونلاين: | http://www.mdpi.com/1424-8220/3/3/43/ |
| الملخص: | Differential pulse voltammetry (DPV) and adsorptive stripping voltammetry (AdSV) at a hanging mercury drop electrode (HMDE) was used for the determination of trace amounts of carcinogenic nitrobiphenyls, namely 2-nitrobiphenyl (2-NBP), 3-nitrobiphenyl (3-NBP) and 4-nitrobiphenyl (4-NBP) within the concentration range from 2.10-8 to 1.10-5 mol⋅L-1 for DPV and from 2.10-9 to 1.10-7 mol⋅L-1 for AdSV using a Britton-Robinson buffer – methanol (1:1) mixture with resulting pH 12 as a base electrolyte. The practical applicability of newly developed methods was verified using model samples of drinking and river water and liquid-liquid extraction for a preliminary separation and preconcentration. |
|---|---|
| تدمد: | 1424-8220 |