Comparison of Different Dosimetric Indices for Volumetric Arc Modulated Treatment Planning Using 2 Different Treatment Planning Systems: A Feasibility Study for Total Body Irradiation

Objective: To compare two different treatment planning systems (TPSs) for plan quality and to check the feasibility of VMAT TBI on the Elekta linear accelerator ‘Versa HD’. Methods: Ten clinical VMAT TBI plans were replanned in Monaco TPS version 5 for retrospective dosimetric analysis and to check...

وصف كامل

التفاصيل البيبلوغرافية
الحاوية / القاعدة:Asian Pacific Journal of Cancer Care
المؤلفون الرئيسيون: Reena Kumari, Budhi Singh Yadav, Pankaj Kumar
التنسيق: مقال
اللغة:الإنجليزية
منشور في: West Asia Organization for Cancer Prevention 2023-07-01
الموضوعات:
الوصول للمادة أونلاين:http://www.waocp.com/journal/index.php/apjcc/article/view/1097
_version_ 1849988371033620480
author Reena Kumari
Budhi Singh Yadav
Pankaj Kumar
author_facet Reena Kumari
Budhi Singh Yadav
Pankaj Kumar
author_sort Reena Kumari
collection DOAJ
container_title Asian Pacific Journal of Cancer Care
description Objective: To compare two different treatment planning systems (TPSs) for plan quality and to check the feasibility of VMAT TBI on the Elekta linear accelerator ‘Versa HD’. Methods: Ten clinical VMAT TBI plans were replanned in Monaco TPS version 5 for retrospective dosimetric analysis and to check the feasibility of treatment delivery in Elekta Versa HD. A dose of 12 Gy in 6 fractions was prescribed to cover 95% PTV. The OARs were the lungs, kidneys, heart, liver, and brain. Dosimetric optimization and calculation were performed in Monaco TPS with a statistical uncertainty of 1%, 2 mm grid. The chosen criterion for plan evaluation and comparison was based on plan quality, dosimetric indices like conformity and homogeneity index, and OAR doses. The comparison of time for optimization, monitor units, and beam on time was also analyzed. Pre-treatment verification using Octavious phantom and gamma analysis using Low’s method with a dose difference of 3% and distance to agreement 3mm criteria was done. The dosimetry of 2 VMAT TBI plans were compared using Wilcoxon signed rank test. Result: The calculated average maximum and mean doses of the PTV were 14.3321Gy and 12.235Gy for eclipse; and 14.428Gy and 12.198Gy for the Monaco system, respectively. The planning time in Monaco was approx. 4.5–6hrs compared to 14–18hrs for Eclipse TPS. Statistically significant difference was observed between the results of 2 TPSs in terms of monitor units, mean doses to lungs, kidneys, heart, and liver and planning time (P < 0.05). However, calculated p values for HI, CI, and brain for both VMAT TBI plans were not statistically different. Conclusion: The study showed the feasibility of VMAT TBI for VERSA HD as it offers faster, better optimization in a single target compared to the Eclipse.
format Article
id doaj-art-ad5fa3de6d624c1eb4eda6aaa9e4c8ab
institution Directory of Open Access Journals
issn 2588-3682
language English
publishDate 2023-07-01
publisher West Asia Organization for Cancer Prevention
record_format Article
spelling doaj-art-ad5fa3de6d624c1eb4eda6aaa9e4c8ab2025-08-20T00:54:44ZengWest Asia Organization for Cancer PreventionAsian Pacific Journal of Cancer Care2588-36822023-07-018348348910.31557/apjcc.2023.8.3.483-4891097Comparison of Different Dosimetric Indices for Volumetric Arc Modulated Treatment Planning Using 2 Different Treatment Planning Systems: A Feasibility Study for Total Body IrradiationReena Kumari0Budhi Singh Yadav1Pankaj Kumar2Department of Applied Sciences, Chitkara University Institute of Engineering and Technology, Chitkara University, Rajpura, Patiala -140401, Punjab, India. Department of Radiotherapy and Oncology, Post Graduate Institute of Medical Education and Research (PGIMER), Chandigarh – 160012, Punjab, India.Department of Radiotherapy and Oncology, Post Graduate Institute of Medical Education and Research (PGIMER), Chandigarh – 160012, Punjab, India.Department of Applied Sciences, Chitkara University Institute of Engineering and Technology, Chitkara University, Rajpura, Patiala -140401, Punjab, IndiaObjective: To compare two different treatment planning systems (TPSs) for plan quality and to check the feasibility of VMAT TBI on the Elekta linear accelerator ‘Versa HD’. Methods: Ten clinical VMAT TBI plans were replanned in Monaco TPS version 5 for retrospective dosimetric analysis and to check the feasibility of treatment delivery in Elekta Versa HD. A dose of 12 Gy in 6 fractions was prescribed to cover 95% PTV. The OARs were the lungs, kidneys, heart, liver, and brain. Dosimetric optimization and calculation were performed in Monaco TPS with a statistical uncertainty of 1%, 2 mm grid. The chosen criterion for plan evaluation and comparison was based on plan quality, dosimetric indices like conformity and homogeneity index, and OAR doses. The comparison of time for optimization, monitor units, and beam on time was also analyzed. Pre-treatment verification using Octavious phantom and gamma analysis using Low’s method with a dose difference of 3% and distance to agreement 3mm criteria was done. The dosimetry of 2 VMAT TBI plans were compared using Wilcoxon signed rank test. Result: The calculated average maximum and mean doses of the PTV were 14.3321Gy and 12.235Gy for eclipse; and 14.428Gy and 12.198Gy for the Monaco system, respectively. The planning time in Monaco was approx. 4.5–6hrs compared to 14–18hrs for Eclipse TPS. Statistically significant difference was observed between the results of 2 TPSs in terms of monitor units, mean doses to lungs, kidneys, heart, and liver and planning time (P < 0.05). However, calculated p values for HI, CI, and brain for both VMAT TBI plans were not statistically different. Conclusion: The study showed the feasibility of VMAT TBI for VERSA HD as it offers faster, better optimization in a single target compared to the Eclipse.http://www.waocp.com/journal/index.php/apjcc/article/view/1097vmat, tbi, oar, mu
spellingShingle Reena Kumari
Budhi Singh Yadav
Pankaj Kumar
Comparison of Different Dosimetric Indices for Volumetric Arc Modulated Treatment Planning Using 2 Different Treatment Planning Systems: A Feasibility Study for Total Body Irradiation
vmat, tbi, oar, mu
title Comparison of Different Dosimetric Indices for Volumetric Arc Modulated Treatment Planning Using 2 Different Treatment Planning Systems: A Feasibility Study for Total Body Irradiation
title_full Comparison of Different Dosimetric Indices for Volumetric Arc Modulated Treatment Planning Using 2 Different Treatment Planning Systems: A Feasibility Study for Total Body Irradiation
title_fullStr Comparison of Different Dosimetric Indices for Volumetric Arc Modulated Treatment Planning Using 2 Different Treatment Planning Systems: A Feasibility Study for Total Body Irradiation
title_full_unstemmed Comparison of Different Dosimetric Indices for Volumetric Arc Modulated Treatment Planning Using 2 Different Treatment Planning Systems: A Feasibility Study for Total Body Irradiation
title_short Comparison of Different Dosimetric Indices for Volumetric Arc Modulated Treatment Planning Using 2 Different Treatment Planning Systems: A Feasibility Study for Total Body Irradiation
title_sort comparison of different dosimetric indices for volumetric arc modulated treatment planning using 2 different treatment planning systems a feasibility study for total body irradiation
topic vmat, tbi, oar, mu
url http://www.waocp.com/journal/index.php/apjcc/article/view/1097
work_keys_str_mv AT reenakumari comparisonofdifferentdosimetricindicesforvolumetricarcmodulatedtreatmentplanningusing2differenttreatmentplanningsystemsafeasibilitystudyfortotalbodyirradiation
AT budhisinghyadav comparisonofdifferentdosimetricindicesforvolumetricarcmodulatedtreatmentplanningusing2differenttreatmentplanningsystemsafeasibilitystudyfortotalbodyirradiation
AT pankajkumar comparisonofdifferentdosimetricindicesforvolumetricarcmodulatedtreatmentplanningusing2differenttreatmentplanningsystemsafeasibilitystudyfortotalbodyirradiation