Surface Morphology and Histopathological Aspects of Metallic Used Cardiovascular CoCr Stents

In this study, CoCr cardiovascular metallic stents with the same chemical composition as well as the adjacent tissues were analyzed. Recovered stents and adjacent tissues were characterized by surface morphology techniques like light microscopy (LM), scanning electron microscopy coupled with energy...

詳細記述

書誌詳細
出版年:Metals
主要な著者: George-Razvan Pandelea-Dobrovicescu, Mariana Prodana, Florentina Golgovici, Daniela Ionita, Maria Sajin, Ioana Demetrescu
フォーマット: 論文
言語:英語
出版事項: MDPI AG 2020-08-01
主題:
オンライン・アクセス:https://www.mdpi.com/2075-4701/10/9/1112
その他の書誌記述
要約:In this study, CoCr cardiovascular metallic stents with the same chemical composition as well as the adjacent tissues were analyzed. Recovered stents and adjacent tissues were characterized by surface morphology techniques like light microscopy (LM), scanning electron microscopy coupled with energy dispersive X-ray spectroscopy (SEM-EDS) and transmission electron microscopy (TEM). The presence of metals in the tissue from around explanted stent was also confirmed by TEM method. The released metal ions were measured using inductively coupled plasma mass spectrometry (ICP-MS). Histopathological analysis was used to highlight host tissue response to the implanted stents. EDX analysis revealed that 21 stents had the same composition. The quantities of Cr ions detected in the tissues were between 5 and 14 ppb and for Co ions were between 4 and 10 ppb. The tissue sections near the layers that generated large quantities of ions showed restenosis development and the deposits of surface calcifications were highlighted. The heavy metal levels observed in this study due to corrosion are too low to cause a systemic toxic response.
ISSN:2075-4701