Effective patient dose during neuroradiological C-arm CT procedures

PURPOSEWe aimed to determine the effective dose given to patients during neuroradiological C-arm computed tomography (CT) procedures.MATERIALS AND METHODSMATERIALS AND METHODSMeasurements were performed on 48 patients using a dose-area product (DAP) meter. A PC-based Monte Carlo program (PCXMC, STUK...

Full description

Bibliographic Details
Published in:Diagnostic and Interventional Radiology
Main Authors: Mei Bai, Xianghua Liu, Bin Liu
Format: Article
Language:English
Published: Turkish Society of Radiology 2013-01-01
Online Access: http://www.dirjournal.org/archives/archive-detail/article-preview/effective-patient-dose-during-neuroradiological-c-/56786
_version_ 1848651505810276352
author Mei Bai,
Xianghua Liu,
Bin Liu
author_facet Mei Bai,
Xianghua Liu,
Bin Liu
author_sort Mei Bai,
collection DOAJ
container_title Diagnostic and Interventional Radiology
description PURPOSEWe aimed to determine the effective dose given to patients during neuroradiological C-arm computed tomography (CT) procedures.MATERIALS AND METHODSMATERIALS AND METHODSMeasurements were performed on 48 patients using a dose-area product (DAP) meter. A PC-based Monte Carlo program (PCXMC, STUK, Helsinki, Finland) was used to calculate the effective dose from the DAP values of each patient. Organ doses were also measured with thermoluminescent dosimeters (TLDs) using a human-shaped phantom. RESULTSThe difference between the organ doses measured using TLDs and PCXMC was not significant (P > 0.05). The mean DAP for 48 patients was 9.41±2.50 Gy·cm2. ; the mean effective dose for all procedures was 0.30±0.08 mSv. The coefficient for the correlation (R 2 2 ) between the DAP and the effective dose was 0.97. The conversion factor between the effective dose and DAP was 0.030–0.035 mSv·Gy-1 ·cm-2 .CONCLUSIONDAP can be used as a dose indicator to calculate the organ dose and effective dose of patients based on Monte Carlo simulation. This method can provide important information on the absorbed dose and enhance the radiation protection of patients during C-arm CT procedures.
format Article
id doaj-e61e425363f34d718e42a2c7118b2bdf
institution Directory of Open Access Journals
issn 1305-3825
1305-3612
language English
publishDate 2013-01-01
publisher Turkish Society of Radiology
record_format Article
spelling doaj-e61e425363f34d718e42a2c7118b2bdf2025-11-03T00:44:52ZengTurkish Society of RadiologyDiagnostic and Interventional Radiology1305-38251305-36122013-01-01191293210.4261/1305-3825.DIR.5852-12.013049054Effective patient dose during neuroradiological C-arm CT proceduresMei Bai,0Xianghua Liu,1Bin Liu2 Department of Biomedical Engineering, Xuanwu Hospital of Capital Medical University, Beijing, China Department of Biomedical Engineering, Xuanwu Hospital of Capital Medical University, Beijing, China Department of Biomedical Engineering, Xuanwu Hospital of Capital Medical University, Beijing, China PURPOSEWe aimed to determine the effective dose given to patients during neuroradiological C-arm computed tomography (CT) procedures.MATERIALS AND METHODSMATERIALS AND METHODSMeasurements were performed on 48 patients using a dose-area product (DAP) meter. A PC-based Monte Carlo program (PCXMC, STUK, Helsinki, Finland) was used to calculate the effective dose from the DAP values of each patient. Organ doses were also measured with thermoluminescent dosimeters (TLDs) using a human-shaped phantom. RESULTSThe difference between the organ doses measured using TLDs and PCXMC was not significant (P > 0.05). The mean DAP for 48 patients was 9.41±2.50 Gy·cm2. ; the mean effective dose for all procedures was 0.30±0.08 mSv. The coefficient for the correlation (R 2 2 ) between the DAP and the effective dose was 0.97. The conversion factor between the effective dose and DAP was 0.030–0.035 mSv·Gy-1 ·cm-2 .CONCLUSIONDAP can be used as a dose indicator to calculate the organ dose and effective dose of patients based on Monte Carlo simulation. This method can provide important information on the absorbed dose and enhance the radiation protection of patients during C-arm CT procedures. http://www.dirjournal.org/archives/archive-detail/article-preview/effective-patient-dose-during-neuroradiological-c-/56786
spellingShingle Mei Bai,
Xianghua Liu,
Bin Liu
Effective patient dose during neuroradiological C-arm CT procedures
title Effective patient dose during neuroradiological C-arm CT procedures
title_full Effective patient dose during neuroradiological C-arm CT procedures
title_fullStr Effective patient dose during neuroradiological C-arm CT procedures
title_full_unstemmed Effective patient dose during neuroradiological C-arm CT procedures
title_short Effective patient dose during neuroradiological C-arm CT procedures
title_sort effective patient dose during neuroradiological c arm ct procedures
url http://www.dirjournal.org/archives/archive-detail/article-preview/effective-patient-dose-during-neuroradiological-c-/56786
work_keys_str_mv AT meibai effectivepatientdoseduringneuroradiologicalcarmctprocedures
AT xianghualiu effectivepatientdoseduringneuroradiologicalcarmctprocedures
AT binliu effectivepatientdoseduringneuroradiologicalcarmctprocedures