A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear Accelerators

Introduction: To study the effect of the International Atomic Energy Agency (IAEA) TRS-483 recommended beam quality correction factor in reference dosimetry and to examine the recommended field output correction factor for relative dosimetry of 6-MV flattening filter free (FFF) small fields, used in...

Full description

Bibliographic Details
Main Authors: Gopinath Mamballikalam, Senthil kumar, Clinto C.O, Jaon bos R.C., Ahamed Basith P.M.
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2021-05-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:https://ijmp.mums.ac.ir/article_16003_fca9f029e208ec3da2954df57f0ccb6e.pdf
id doaj-01cd303cc9494813ae2216e75c4f1367
record_format Article
spelling doaj-01cd303cc9494813ae2216e75c4f13672021-05-10T07:10:40ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722345-36722021-05-0118317818610.22038/ijmp.2020.46236.172216003A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear AcceleratorsGopinath Mamballikalam0Senthil kumar1Clinto C.O2Jaon bos R.C.3Ahamed Basith P.M.41. R&D, Bharathiar University, Coimbatore, Tamilnadu, India 3. Aster Medcity, Kochi, Kerala, IndiaRajaji Hospital & Madurai Medical College, Madurai, Tamilnadu, IndiaAster Medcity, Kochi, Kerala, IndiaAster Medcity, Kochi, Kerala, IndiaAster Medcity, Kochi, Kerala, IndiaIntroduction: To study the effect of the International Atomic Energy Agency (IAEA) TRS-483 recommended beam quality correction factor in reference dosimetry and to examine the recommended field output correction factor for relative dosimetry of 6-MV flattening filter free (FFF) small fields, used in a Varian TrueBeam linear accelerator (LINAC). Material and Methods: The beam quality and field output correction for 6-MV FFF beams were adopted from the TRS-483 protocol. Monte Carlo (MC) simulation of the output factor was performed using the PENELOPE-based PRIMO software and compared with the TRS-483 corrected output factors. Two analytical anisotropic algorithm (AAA) models in the EclipseTM treatment planning system (TPS) were created; one with an output factor taken as the ratio of meter readings and one with an output factor obtained by multiplying the TRS-483 correction factor by the ratio of meter readings. Besides, box field and dynamic conformal arc (DCA) plans were created for both AAA models for verification and validation. The patient-specific quality assurances (QA) for ten different targets were performed, and deviations between the measured and TPS-calculated point doses in both models were examined. Results: Separate beam quality correction factors for FFF beams in the TRS-483 protocol only resulted in an improvement of 0.1% in reference dosimetry. The TRS-483 corrected output factor was in a better agreement with the MC-calculated output factor. For a patient-specific QA of DCA plans, the output factor-corrected AAA dose calculation algorithm showed a better agreement between the measured and simulated doses. Also, there was a smaller deviation (1.2%) for the smallest target of 0.23 cc (8 mm equivalent sphere diameter) used in this study. Conclusion:The field output factors for the LINAC small beams can be improved by incorporating the TRS-483 correction factors. However, the extent of improvement that can be expected depends on the source model of the calculation algorithm and how these well-generalized corrections are suitable for user beams and detectors.https://ijmp.mums.ac.ir/article_16003_fca9f029e208ec3da2954df57f0ccb6e.pdffff beams trs483 linear accelerator small field dosimetry
collection DOAJ
language English
format Article
sources DOAJ
author Gopinath Mamballikalam
Senthil kumar
Clinto C.O
Jaon bos R.C.
Ahamed Basith P.M.
spellingShingle Gopinath Mamballikalam
Senthil kumar
Clinto C.O
Jaon bos R.C.
Ahamed Basith P.M.
A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear Accelerators
Iranian Journal of Medical Physics
fff beams trs
483 linear accelerator small field dosimetry
author_facet Gopinath Mamballikalam
Senthil kumar
Clinto C.O
Jaon bos R.C.
Ahamed Basith P.M.
author_sort Gopinath Mamballikalam
title A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear Accelerators
title_short A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear Accelerators
title_full A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear Accelerators
title_fullStr A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear Accelerators
title_full_unstemmed A Study of the Impact of TRS-483 Recommended Correction Factors for Dosimetry of Small Fields in Flattening Filter Free Beams used in TrueBeam Linear Accelerators
title_sort study of the impact of trs-483 recommended correction factors for dosimetry of small fields in flattening filter free beams used in truebeam linear accelerators
publisher Mashhad University of Medical Sciences
series Iranian Journal of Medical Physics
issn 2345-3672
2345-3672
publishDate 2021-05-01
description Introduction: To study the effect of the International Atomic Energy Agency (IAEA) TRS-483 recommended beam quality correction factor in reference dosimetry and to examine the recommended field output correction factor for relative dosimetry of 6-MV flattening filter free (FFF) small fields, used in a Varian TrueBeam linear accelerator (LINAC). Material and Methods: The beam quality and field output correction for 6-MV FFF beams were adopted from the TRS-483 protocol. Monte Carlo (MC) simulation of the output factor was performed using the PENELOPE-based PRIMO software and compared with the TRS-483 corrected output factors. Two analytical anisotropic algorithm (AAA) models in the EclipseTM treatment planning system (TPS) were created; one with an output factor taken as the ratio of meter readings and one with an output factor obtained by multiplying the TRS-483 correction factor by the ratio of meter readings. Besides, box field and dynamic conformal arc (DCA) plans were created for both AAA models for verification and validation. The patient-specific quality assurances (QA) for ten different targets were performed, and deviations between the measured and TPS-calculated point doses in both models were examined. Results: Separate beam quality correction factors for FFF beams in the TRS-483 protocol only resulted in an improvement of 0.1% in reference dosimetry. The TRS-483 corrected output factor was in a better agreement with the MC-calculated output factor. For a patient-specific QA of DCA plans, the output factor-corrected AAA dose calculation algorithm showed a better agreement between the measured and simulated doses. Also, there was a smaller deviation (1.2%) for the smallest target of 0.23 cc (8 mm equivalent sphere diameter) used in this study. Conclusion:The field output factors for the LINAC small beams can be improved by incorporating the TRS-483 correction factors. However, the extent of improvement that can be expected depends on the source model of the calculation algorithm and how these well-generalized corrections are suitable for user beams and detectors.
topic fff beams trs
483 linear accelerator small field dosimetry
url https://ijmp.mums.ac.ir/article_16003_fca9f029e208ec3da2954df57f0ccb6e.pdf
work_keys_str_mv AT gopinathmamballikalam astudyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT senthilkumar astudyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT clintoco astudyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT jaonbosrc astudyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT ahamedbasithpm astudyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT gopinathmamballikalam studyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT senthilkumar studyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT clintoco studyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT jaonbosrc studyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
AT ahamedbasithpm studyoftheimpactoftrs483recommendedcorrectionfactorsfordosimetryofsmallfieldsinflatteningfilterfreebeamsusedintruebeamlinearaccelerators
_version_ 1721453447072448512