Verification of an interaction model of an ultrasonic oscillatory system with periodontal tissues

Verification of an interaction model of an ultrasonic oscillatory system with biological tissues which was developed in COMSOL Multiphysics was carried out. It was shown that calculation results in COMSOL Multiphysics obtained using the “Finer” grid (the ratio of the grid step to a minimum transvers...

Full description

Bibliographic Details
Main Authors: V. A. Karpuhin, N. V. Vishneva
Format: Article
Language:Russian
Published: MGTU im. N.È. Baumana 2014-01-01
Series:Nauka i Obrazovanie
Subjects:
Online Access:http://technomag.edu.ru/jour/article/view/454
Description
Summary:Verification of an interaction model of an ultrasonic oscillatory system with biological tissues which was developed in COMSOL Multiphysics was carried out. It was shown that calculation results in COMSOL Multiphysics obtained using the “Finer” grid (the ratio of the grid step to a minimum transversal section area of the model ≤ 0.3 mm-1) best of all qualitatively and quantitatively corresponded to practical results. The average relative error of the obtained results in comparison with the experimental ones did not exceed 4.0%. Influence of geometrical parameters (thickness of load) on electrical admittance of the ultrasonic oscillatory system interacting with biological tissues was investigated. It was shown that increase in thickness of load within the range from 0 to 95 mm led to decrease in calculated values of natural resonance frequency of longitudinal fluctuations and electrical admittance from 26,58 to 26,35 kHz and from 0,86 to 0,44 mS.
ISSN:1994-0408