“Add-on pacing lead”: An effective and safe alternative to lead replacement in ICD pacing failure
Despite recent advances in implantable cardioverter defibrillator (ICD) technology, the long-term reliability of ICD leads remains a significant problem. Lead failures constitute a major risk for patients with an implantable cardioverter defibrillator. There is no clear consensus on treatment strate...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2012-07-01
|
Series: | Indian Heart Journal |
Subjects: | |
Online Access: | http://www.sciencedirect.com/science/article/pii/S0019483212000417 |
Summary: | Despite recent advances in implantable cardioverter defibrillator (ICD) technology, the long-term reliability of ICD leads remains a significant problem. Lead failures constitute a major risk for patients with an implantable cardioverter defibrillator. There is no clear consensus on treatment strategy of ICD lead failure and decision should be individualized. We report a pacing-dependent elderly male with ICD lead pacing failure secondary to insulation break resulting in recurrent syncope. We emphasize the technique of “Add-on pacing lead implantation” could be an effective and alternative to ICD lead replacement. |
---|---|