Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument

This research in determining x-ray effective energy in linear accelerator (Linac) radiotherapy device was done using copper plates at different thicknesses by determining the Half Value Layer (HVL) for each energy. The purpose of this research is to know the amount of effective energy used in radiot...

Full description

Bibliographic Details
Main Authors: Fauziah Fauziah, Bualkar Abdullah
Format: Article
Language:Indonesian
Published: Institut Agama Islam Negeri Raden Intan Lampung 2018-04-01
Series:Jurnal Ilmiah Pendidikan Fisika Al-Biruni
Subjects:
Online Access:http://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/2336
id doaj-221407e658f64ef4810ded8f757b9e66
record_format Article
spelling doaj-221407e658f64ef4810ded8f757b9e662020-11-25T00:35:19ZindInstitut Agama Islam Negeri Raden Intan LampungJurnal Ilmiah Pendidikan Fisika Al-Biruni2303-18322503-023X2018-04-01719910410.24042/jipfalbiruni.v7i1.23361893Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy InstrumentFauziah Fauziah0Bualkar Abdullah1Physics Department, Hasanuddin UniversityPhysics Departement, Hasanuddin UniversityThis research in determining x-ray effective energy in linear accelerator (Linac) radiotherapy device was done using copper plates at different thicknesses by determining the Half Value Layer (HVL) for each energy. The purpose of this research is to know the amount of effective energy used in radiotherapy linear accelerator. The results showed that the energy of 6 MeV obtained 2,714 MeV of effective energy at HVL 4,500 mm and at the energy of 10 MeV obtained 4,231 MeV of effective energy at HVL 4,500 mm.http://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/2336Copper plates, Effective energy, Half value layer, Linear accelerator
collection DOAJ
language Indonesian
format Article
sources DOAJ
author Fauziah Fauziah
Bualkar Abdullah
spellingShingle Fauziah Fauziah
Bualkar Abdullah
Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument
Jurnal Ilmiah Pendidikan Fisika Al-Biruni
Copper plates, Effective energy, Half value layer, Linear accelerator
author_facet Fauziah Fauziah
Bualkar Abdullah
author_sort Fauziah Fauziah
title Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument
title_short Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument
title_full Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument
title_fullStr Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument
title_full_unstemmed Determining X-Ray Effective Energy in Linear Accelerator (LINAC) Radiotherapy Instrument
title_sort determining x-ray effective energy in linear accelerator (linac) radiotherapy instrument
publisher Institut Agama Islam Negeri Raden Intan Lampung
series Jurnal Ilmiah Pendidikan Fisika Al-Biruni
issn 2303-1832
2503-023X
publishDate 2018-04-01
description This research in determining x-ray effective energy in linear accelerator (Linac) radiotherapy device was done using copper plates at different thicknesses by determining the Half Value Layer (HVL) for each energy. The purpose of this research is to know the amount of effective energy used in radiotherapy linear accelerator. The results showed that the energy of 6 MeV obtained 2,714 MeV of effective energy at HVL 4,500 mm and at the energy of 10 MeV obtained 4,231 MeV of effective energy at HVL 4,500 mm.
topic Copper plates, Effective energy, Half value layer, Linear accelerator
url http://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/2336
work_keys_str_mv AT fauziahfauziah determiningxrayeffectiveenergyinlinearacceleratorlinacradiotherapyinstrument
AT bualkarabdullah determiningxrayeffectiveenergyinlinearacceleratorlinacradiotherapyinstrument
_version_ 1725308989181263872