Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth

Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods....

Full description

Bibliographic Details
Main Authors: Jinhai Huang, Weicong Lu, Giacomo Savini, Hao Chen, Chengfang Wang, Xinxin Yu, Fangjun Bao, Qinmei Wang
Format: Article
Language:English
Published: Hindawi Limited 2016-01-01
Series:Journal of Ophthalmology
Online Access:http://dx.doi.org/10.1155/2016/6347236
id doaj-23ba223facb84edb95a5f258700e0976
record_format Article
spelling doaj-23ba223facb84edb95a5f258700e09762020-11-24T23:48:03ZengHindawi LimitedJournal of Ophthalmology2090-004X2090-00582016-01-01201610.1155/2016/63472366347236Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber DepthJinhai Huang0Weicong Lu1Giacomo Savini2Hao Chen3Chengfang Wang4Xinxin Yu5Fangjun Bao6Qinmei Wang7School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaG.B. Bietti Foundation IRCCS, Rome, ItalySchool of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaSchool of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, ChinaPurpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59±27.37 μm, 3.70±0.30 mm, and 3.16±0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14±26.61 μm for CCT, 3.71±0.29 mm for ACD, and 3.17±0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45±6.07 μm, P<0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.http://dx.doi.org/10.1155/2016/6347236
collection DOAJ
language English
format Article
sources DOAJ
author Jinhai Huang
Weicong Lu
Giacomo Savini
Hao Chen
Chengfang Wang
Xinxin Yu
Fangjun Bao
Qinmei Wang
spellingShingle Jinhai Huang
Weicong Lu
Giacomo Savini
Hao Chen
Chengfang Wang
Xinxin Yu
Fangjun Bao
Qinmei Wang
Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
Journal of Ophthalmology
author_facet Jinhai Huang
Weicong Lu
Giacomo Savini
Hao Chen
Chengfang Wang
Xinxin Yu
Fangjun Bao
Qinmei Wang
author_sort Jinhai Huang
title Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_short Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_full Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_fullStr Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_full_unstemmed Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth
title_sort comparison between a new optical biometry device and an anterior segment optical coherence tomographer for measuring central corneal thickness and anterior chamber depth
publisher Hindawi Limited
series Journal of Ophthalmology
issn 2090-004X
2090-0058
publishDate 2016-01-01
description Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59±27.37 μm, 3.70±0.30 mm, and 3.16±0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14±26.61 μm for CCT, 3.71±0.29 mm for ACD, and 3.17±0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45±6.07 μm, P<0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.
url http://dx.doi.org/10.1155/2016/6347236
work_keys_str_mv AT jinhaihuang comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT weiconglu comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT giacomosavini comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT haochen comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT chengfangwang comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT xinxinyu comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT fangjunbao comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
AT qinmeiwang comparisonbetweenanewopticalbiometrydeviceandananteriorsegmentopticalcoherencetomographerformeasuringcentralcornealthicknessandanteriorchamberdepth
_version_ 1725487491024158720