Set-up of a method using LC-UV to assay mometasone furoate in pharmaceutical dosage forms
A reliable method using LC-UV to assay mometasone furoate (MF) in creams or nasal sprays using the same chromatographic conditions was set up. Methanol:water 80:20 (v/v) (1.0 mL min-1) was used as mobile phase. MF was detected at 248 nm and analyzed in a concentration range from 1.0 to 20.0 µg mL-1....
Main Authors: | , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Sociedade Brasileira de Química
2012-01-01
|
Series: | Química Nova |
Subjects: | |
Online Access: | http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422012000400030&lng=en&tlng=en |
Summary: | A reliable method using LC-UV to assay mometasone furoate (MF) in creams or nasal sprays using the same chromatographic conditions was set up. Methanol:water 80:20 (v/v) (1.0 mL min-1) was used as mobile phase. MF was detected at 248 nm and analyzed in a concentration range from 1.0 to 20.0 µg mL-1. The method provided acceptable theoretical plates, peak simmetry, peak tailing factor and peak resolution a short run (5 min). The method showed specificity, good linearity (r = 0.9999) and the quantification limit was 0.379 µg mL-1. Furthermore, the method was precise (RSD < 2.0%), accurate (recovery > 97%) and robust. |
---|---|
ISSN: | 1678-7064 |