Validity and Reliability Study of Two Different Portable Electro–Enzymatic Lactate Analysers
Objectives: The purpose of this study was to evaluate the reliabilities of the Biosen C Line (BCL) laboratory and Lactate Scout (LS) hand-held portable lactate analyser within a wide range lactate concentration. LS analyser was compared to the BCL analyser (reference method) to evaluate its validity...
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Turkish Sports Medicine Association
2020-06-01
|
Series: | Spor Hekimligi Dergisi |
Subjects: | |
Online Access: |
https://journalofsportsmedicine.org/eng/full-text-pdf/532/tur
|
id |
doaj-29a85e9b20c649fe83e07be0270431da |
---|---|
record_format |
Article |
spelling |
doaj-29a85e9b20c649fe83e07be0270431da2020-11-25T03:14:12ZengTurkish Sports Medicine AssociationSpor Hekimligi Dergisi1300-05512587-14982020-06-0155213814710.5152/tjsm.2020.170532Validity and Reliability Study of Two Different Portable Electro–Enzymatic Lactate AnalysersHalil İbrahim Kaya0Cem Çetin1Duygu Doğuç Kumbul2Mustafa Onur Serbest3Ali Erdoğan4 Sports Medicine Section, Samsun Training and Research Hospital, University of Health Sciences, Samsun, Turkey Department of Sports Medicine, Faculty of Medicine, Süleyman Demirel University, Isparta, Turkey Department of Medical Biochemistry, Faculty of Medicine, Süleyman Demirel University, Isparta, Turkey Sports Medicine Section, Aydın State Hospital, Aydın, Turkey Sports Medicine Unit, Gloria Sports Arena, Antalya, Turkey Objectives: The purpose of this study was to evaluate the reliabilities of the Biosen C Line (BCL) laboratory and Lactate Scout (LS) hand-held portable lactate analyser within a wide range lactate concentration. LS analyser was compared to the BCL analyser (reference method) to evaluate its validity. Materials and methods: Blood samples were taken during an incremental treadmill test under laboratory conditions. In order to evaluate reliability, 99 blood samples with lactate concentrations changed between 0.8-11.1 mM were measured twice. Results: For the LS analyser, intraclass correlation coefficient (ICC) was 0.994; typical error of measurement (TE) and coefficient of variation (CV) were 0.200 mM and 2.95%, respectively. For the BCL analyser, ICC was 1.000; TE and CV were 0.044 mM and 0.50%, respectively. LS hand-held portable- and BCL laboratory analysers’ reliability results show perfect accordance. In order to evaluate LS analyser reliability, 99 blood samples with lactate concentrations between 0.8-11.1 mM were also measured, giving a high validity coefficient of 0.976. Validity was also separately assessed for the three sample ranges of 5.0 mM, yielding ICC values of 0.862, 0.903 and 0.783, respectively. Comparing with the Bland & Altman method, LS and reference method mean lactic acid concentration difference was -0.097 mM, and limits of agreement -1.342 and 1.147 mM. Conclusion: The results of the study revealed that the LS analyser has high validity, however at concentrations higher than 5.0 mM, its measures are found significantly lower than those of the reference method. https://journalofsportsmedicine.org/eng/full-text-pdf/532/tur lactate analysisvalidityreliability |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Halil İbrahim Kaya Cem Çetin Duygu Doğuç Kumbul Mustafa Onur Serbest Ali Erdoğan |
spellingShingle |
Halil İbrahim Kaya Cem Çetin Duygu Doğuç Kumbul Mustafa Onur Serbest Ali Erdoğan Validity and Reliability Study of Two Different Portable Electro–Enzymatic Lactate Analysers Spor Hekimligi Dergisi lactate analysis validity reliability |
author_facet |
Halil İbrahim Kaya Cem Çetin Duygu Doğuç Kumbul Mustafa Onur Serbest Ali Erdoğan |
author_sort |
Halil İbrahim Kaya |
title |
Validity and Reliability Study of Two Different Portable Electro–Enzymatic Lactate Analysers |
title_short |
Validity and Reliability Study of Two Different Portable Electro–Enzymatic Lactate Analysers |
title_full |
Validity and Reliability Study of Two Different Portable Electro–Enzymatic Lactate Analysers |
title_fullStr |
Validity and Reliability Study of Two Different Portable Electro–Enzymatic Lactate Analysers |
title_full_unstemmed |
Validity and Reliability Study of Two Different Portable Electro–Enzymatic Lactate Analysers |
title_sort |
validity and reliability study of two different portable electro–enzymatic lactate analysers |
publisher |
Turkish Sports Medicine Association |
series |
Spor Hekimligi Dergisi |
issn |
1300-0551 2587-1498 |
publishDate |
2020-06-01 |
description |
Objectives: The purpose of this study was to evaluate the reliabilities of the Biosen C Line (BCL) laboratory and Lactate Scout (LS) hand-held portable lactate analyser within a wide range lactate concentration. LS analyser was compared to the BCL analyser (reference method) to evaluate its validity.
Materials and methods: Blood samples were taken during an incremental treadmill test under laboratory conditions. In order to evaluate reliability, 99 blood samples with lactate concentrations changed between 0.8-11.1 mM were measured twice.
Results: For the LS analyser, intraclass correlation coefficient (ICC) was 0.994; typical error of measurement (TE) and coefficient of variation (CV) were 0.200 mM and 2.95%, respectively. For the BCL analyser, ICC was 1.000; TE and CV were 0.044 mM and 0.50%, respectively. LS hand-held portable- and BCL laboratory analysers’ reliability results show perfect accordance. In order to evaluate LS analyser reliability, 99 blood samples with lactate concentrations between 0.8-11.1 mM were also measured, giving a high validity coefficient of 0.976. Validity was also separately assessed for the three sample ranges of 5.0 mM, yielding ICC values of 0.862, 0.903 and 0.783, respectively. Comparing with the Bland & Altman method, LS and reference method mean lactic acid concentration difference was -0.097 mM, and limits of agreement -1.342 and 1.147 mM.
Conclusion: The results of the study revealed that the LS analyser has high validity, however at concentrations higher than 5.0 mM, its measures are found significantly lower than those of the reference method. |
topic |
lactate analysis validity reliability |
url |
https://journalofsportsmedicine.org/eng/full-text-pdf/532/tur
|
work_keys_str_mv |
AT halilibrahimkaya validityandreliabilitystudyoftwodifferentportableelectroenzymaticlactateanalysers AT cemcetin validityandreliabilitystudyoftwodifferentportableelectroenzymaticlactateanalysers AT duygudoguckumbul validityandreliabilitystudyoftwodifferentportableelectroenzymaticlactateanalysers AT mustafaonurserbest validityandreliabilitystudyoftwodifferentportableelectroenzymaticlactateanalysers AT alierdogan validityandreliabilitystudyoftwodifferentportableelectroenzymaticlactateanalysers |
_version_ |
1724643962246922240 |