Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China

Shuyun Chen,1,2,* Xiaochen Li,1,3,* Zihui Wang,1,* Yumin Zhou,1 Dongxing Zhao,1 Zhuxiang Zhao,1,4 Sha Liu,1 Pixin Ran1 1State Key Laboratory of Respiratory Disease, Guangzhou Institute of Respiratory Diseases, National Clinical Research Center for Respiratory Disease Guangzho...

Full description

Bibliographic Details
Main Authors: Chen S, Li X, Wang Z, Zhou Y, Zhao D, Zhao Z, Liu S, Ran P
Format: Article
Language:English
Published: Dove Medical Press 2021-07-01
Series:International Journal of COPD
Subjects:
Online Access:https://www.dovepress.com/validity-of-the-handheld-expiratory-flowmeter-for-copd-screening-in-th-peer-reviewed-fulltext-article-COPD
id doaj-2b7968971b8a4d99914891df13506282
record_format Article
spelling doaj-2b7968971b8a4d99914891df135062822021-07-11T19:23:06ZengDove Medical PressInternational Journal of COPD1178-20052021-07-01Volume 162039204766727Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of ChinaChen SLi XWang ZZhou YZhao DZhao ZLiu SRan PShuyun Chen,1,2,&ast; Xiaochen Li,1,3,&ast; Zihui Wang,1,&ast; Yumin Zhou,1 Dongxing Zhao,1 Zhuxiang Zhao,1,4 Sha Liu,1 Pixin Ran1 1State Key Laboratory of Respiratory Disease, Guangzhou Institute of Respiratory Diseases, National Clinical Research Center for Respiratory Disease Guangzhou, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, Guangzhou, Guangdong Province, People’s Republic of China; 2Department of Respiratory and Critical Care Medicine, First Affiliated Hospital of Nanchang University, Nanchang, Jiangxi Province, People’s Republic of China; 3The People’s Hospital of Hubei Province, Wuhan, Hubei Province, People’s Republic of China; 4The First People’s Hospital of Guangzhou City, Guangzhou, Guangdong Province, People’s Republic of China&ast;These authors contributed equally to this workCorrespondence: Pixin RanState Key Laboratory of Respiratory Disease, Guangzhou Institute of Respiratory Diseases, National Clinical Research Center for Respiratory Disease Guangzhou, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, No. 195 Dongfeng Xi Road, Guangzhou, 510000, Guangdong Province, People’s Republic of ChinaTel +86 2083205187Email pxran@gzhmu.edu.cnPurpose: The use of simple and affordable screening tools for chronic obstructive pulmonary disease (COPD) is limited. We aimed to assess the validity of a handheld expiratory flowmeter (Vitalograph Ltd., COPD-6®, Ireland) for COPD screening in Chinese primary care settings.Methods: In our cross-sectional study, subjects were randomly selected in eight primary care settings. Tests with the handheld expiratory flowmeter and the conventional spirometry were sequentially performed on all participants. The correlation between the handheld expiratory flowmeter and the conventional spirometry was determined. Validity was determined by the area under the receiver operator characteristic curve (AUC) of the forced expiratory volume in one second (FEV1)/forced expiratory volume in six seconds (FEV6) that used to detect airway obstruction. The sensitivity, specificity, predictive values, and likelihood ratio were calculated according to different FEV1/FEV6 cut-off points.Results: A total of 229 subjects (15.4%) were diagnosed with airflow limitation by conventional spirometry. FEV1, FEV6, and FEV1/FEV6 measured by the handheld expiratory flowmeter were correlated with FEV1, FVC, and FEV1/FVC measured by the conventional spirometry (r=0.889, 0.835 and 0.647, p< 0.001), respectively. AUC of the FEV1/FEV6 to determine airflow obstruction was 0.857 (95% CI: 0.826 to 0.888). No significant difference of AUC was observed between the symptomatic group and the asymptomatic group (AUC=0.869 vs 0.843, P=0.425). A similar phenomenon was found in the AUC of smokers and never-smokers (AUC=0.862 vs 0.840; P=0.515). The cut-off point for FEV1/FEV6 was 0.77 and the corresponding sensitivity and specificity were 71.2% and 89.8%, respectively.Conclusion: The handheld expiratory flowmeter might be used as a screening device for COPD in Chinese primary care settings.Keywords: COPD screening, airflow limitation, FEV1, FEV6, lung functionhttps://www.dovepress.com/validity-of-the-handheld-expiratory-flowmeter-for-copd-screening-in-th-peer-reviewed-fulltext-article-COPDcopd screeningairflow limitationfev1fev6lung function
collection DOAJ
language English
format Article
sources DOAJ
author Chen S
Li X
Wang Z
Zhou Y
Zhao D
Zhao Z
Liu S
Ran P
spellingShingle Chen S
Li X
Wang Z
Zhou Y
Zhao D
Zhao Z
Liu S
Ran P
Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China
International Journal of COPD
copd screening
airflow limitation
fev1
fev6
lung function
author_facet Chen S
Li X
Wang Z
Zhou Y
Zhao D
Zhao Z
Liu S
Ran P
author_sort Chen S
title Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China
title_short Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China
title_full Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China
title_fullStr Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China
title_full_unstemmed Validity of the Handheld Expiratory Flowmeter for COPD Screening in the Primary Care Setting of China
title_sort validity of the handheld expiratory flowmeter for copd screening in the primary care setting of china
publisher Dove Medical Press
series International Journal of COPD
issn 1178-2005
publishDate 2021-07-01
description Shuyun Chen,1,2,&ast; Xiaochen Li,1,3,&ast; Zihui Wang,1,&ast; Yumin Zhou,1 Dongxing Zhao,1 Zhuxiang Zhao,1,4 Sha Liu,1 Pixin Ran1 1State Key Laboratory of Respiratory Disease, Guangzhou Institute of Respiratory Diseases, National Clinical Research Center for Respiratory Disease Guangzhou, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, Guangzhou, Guangdong Province, People’s Republic of China; 2Department of Respiratory and Critical Care Medicine, First Affiliated Hospital of Nanchang University, Nanchang, Jiangxi Province, People’s Republic of China; 3The People’s Hospital of Hubei Province, Wuhan, Hubei Province, People’s Republic of China; 4The First People’s Hospital of Guangzhou City, Guangzhou, Guangdong Province, People’s Republic of China&ast;These authors contributed equally to this workCorrespondence: Pixin RanState Key Laboratory of Respiratory Disease, Guangzhou Institute of Respiratory Diseases, National Clinical Research Center for Respiratory Disease Guangzhou, First Affiliated Hospital of Guangzhou Medical University, Guangzhou Medical University, No. 195 Dongfeng Xi Road, Guangzhou, 510000, Guangdong Province, People’s Republic of ChinaTel +86 2083205187Email pxran@gzhmu.edu.cnPurpose: The use of simple and affordable screening tools for chronic obstructive pulmonary disease (COPD) is limited. We aimed to assess the validity of a handheld expiratory flowmeter (Vitalograph Ltd., COPD-6®, Ireland) for COPD screening in Chinese primary care settings.Methods: In our cross-sectional study, subjects were randomly selected in eight primary care settings. Tests with the handheld expiratory flowmeter and the conventional spirometry were sequentially performed on all participants. The correlation between the handheld expiratory flowmeter and the conventional spirometry was determined. Validity was determined by the area under the receiver operator characteristic curve (AUC) of the forced expiratory volume in one second (FEV1)/forced expiratory volume in six seconds (FEV6) that used to detect airway obstruction. The sensitivity, specificity, predictive values, and likelihood ratio were calculated according to different FEV1/FEV6 cut-off points.Results: A total of 229 subjects (15.4%) were diagnosed with airflow limitation by conventional spirometry. FEV1, FEV6, and FEV1/FEV6 measured by the handheld expiratory flowmeter were correlated with FEV1, FVC, and FEV1/FVC measured by the conventional spirometry (r=0.889, 0.835 and 0.647, p< 0.001), respectively. AUC of the FEV1/FEV6 to determine airflow obstruction was 0.857 (95% CI: 0.826 to 0.888). No significant difference of AUC was observed between the symptomatic group and the asymptomatic group (AUC=0.869 vs 0.843, P=0.425). A similar phenomenon was found in the AUC of smokers and never-smokers (AUC=0.862 vs 0.840; P=0.515). The cut-off point for FEV1/FEV6 was 0.77 and the corresponding sensitivity and specificity were 71.2% and 89.8%, respectively.Conclusion: The handheld expiratory flowmeter might be used as a screening device for COPD in Chinese primary care settings.Keywords: COPD screening, airflow limitation, FEV1, FEV6, lung function
topic copd screening
airflow limitation
fev1
fev6
lung function
url https://www.dovepress.com/validity-of-the-handheld-expiratory-flowmeter-for-copd-screening-in-th-peer-reviewed-fulltext-article-COPD
work_keys_str_mv AT chens validityofthehandheldexpiratoryflowmeterforcopdscreeningintheprimarycaresettingofchina
AT lix validityofthehandheldexpiratoryflowmeterforcopdscreeningintheprimarycaresettingofchina
AT wangz validityofthehandheldexpiratoryflowmeterforcopdscreeningintheprimarycaresettingofchina
AT zhouy validityofthehandheldexpiratoryflowmeterforcopdscreeningintheprimarycaresettingofchina
AT zhaod validityofthehandheldexpiratoryflowmeterforcopdscreeningintheprimarycaresettingofchina
AT zhaoz validityofthehandheldexpiratoryflowmeterforcopdscreeningintheprimarycaresettingofchina
AT lius validityofthehandheldexpiratoryflowmeterforcopdscreeningintheprimarycaresettingofchina
AT ranp validityofthehandheldexpiratoryflowmeterforcopdscreeningintheprimarycaresettingofchina
_version_ 1721308199755186176