P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE

Introduction: Fully automated cuff-based devices have been developed for the assessment of arterial stiffness via pulse wave velocity (PWV) measurement, such as the VaSera device (Fukuda Denshi). To date, measurements were confined to the heart-to-ankle segment, yielding PWV and stiffness indices th...

Full description

Bibliographic Details
Main Authors: Francisco Londoño*, Daime Campos, Shigeo Horinaka, Julio Chirinos, Patrick Segers
Format: Article
Language:English
Published: Atlantis Press 2015-11-01
Series:Artery Research
Online Access:https://www.atlantis-press.com/article/125930628/view
id doaj-2cf760d259094fa99542905a8d855c8b
record_format Article
spelling doaj-2cf760d259094fa99542905a8d855c8b2020-11-25T03:27:43ZengAtlantis PressArtery Research 1876-44012015-11-011210.1016/j.artres.2015.10.278P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICEFrancisco Londoño*Daime CamposShigeo HorinakaJulio ChirinosPatrick SegersIntroduction: Fully automated cuff-based devices have been developed for the assessment of arterial stiffness via pulse wave velocity (PWV) measurement, such as the VaSera device (Fukuda Denshi). To date, measurements were confined to the heart-to-ankle segment, yielding PWV and stiffness indices that are not easily inked to carotid-femoral PWV (cf-PWV), the presumed reference for measurement of aortic PWV. Methods: We performed initial tests (N=14, 9 males, mean age 27.4±3.3, BMI 23.8±3.4) using a novel thigh-cuff prototype that can be used as a substitute for the ankle cuffs in the VaSera device. Extracted data included heart-thigh (ht-PWV) and heart-ankle (ha-PWV). cf-PWV was obtained using ultrasound (GE Vivid 7) on the right side. Results: Measurements were successfully obtained for all subjects. cf-PWV was 5.32 ± 0.43 m/s. ha-PWV was 6.35±0.49 m/s, and was significantly higher than cf-PWV (paired t-test; P<0.001). ht-PWV, on the other hand, was 5.51±0.50 m/s and was not significantly different from cf-PWV. Bland-Altman analysis demonstrated a non-significant bias of 0.19±0.54 m/s of ht-PWV with respect to cf-PWV. Conclusion: We conclude that, in this small-sized young and healthy population, fully automated measurement of heart-thigh PWV is straightforward and easy. Measured values were not different from carotid-femoral PWV. Further research is warranted to confirm these findings in a larger population spanning a large age range and cardiovascular risk profiles.https://www.atlantis-press.com/article/125930628/view
collection DOAJ
language English
format Article
sources DOAJ
author Francisco Londoño*
Daime Campos
Shigeo Horinaka
Julio Chirinos
Patrick Segers
spellingShingle Francisco Londoño*
Daime Campos
Shigeo Horinaka
Julio Chirinos
Patrick Segers
P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE
Artery Research
author_facet Francisco Londoño*
Daime Campos
Shigeo Horinaka
Julio Chirinos
Patrick Segers
author_sort Francisco Londoño*
title P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE
title_short P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE
title_full P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE
title_fullStr P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE
title_full_unstemmed P5.14 THIGH-CUFF BASED MEASUREMENT OF AORTIC PULSE WAVE VELOCITY: INITIAL TESTING OF A NOVEL VASERA PROTOTYPE DEVICE
title_sort p5.14 thigh-cuff based measurement of aortic pulse wave velocity: initial testing of a novel vasera prototype device
publisher Atlantis Press
series Artery Research
issn 1876-4401
publishDate 2015-11-01
description Introduction: Fully automated cuff-based devices have been developed for the assessment of arterial stiffness via pulse wave velocity (PWV) measurement, such as the VaSera device (Fukuda Denshi). To date, measurements were confined to the heart-to-ankle segment, yielding PWV and stiffness indices that are not easily inked to carotid-femoral PWV (cf-PWV), the presumed reference for measurement of aortic PWV. Methods: We performed initial tests (N=14, 9 males, mean age 27.4±3.3, BMI 23.8±3.4) using a novel thigh-cuff prototype that can be used as a substitute for the ankle cuffs in the VaSera device. Extracted data included heart-thigh (ht-PWV) and heart-ankle (ha-PWV). cf-PWV was obtained using ultrasound (GE Vivid 7) on the right side. Results: Measurements were successfully obtained for all subjects. cf-PWV was 5.32 ± 0.43 m/s. ha-PWV was 6.35±0.49 m/s, and was significantly higher than cf-PWV (paired t-test; P<0.001). ht-PWV, on the other hand, was 5.51±0.50 m/s and was not significantly different from cf-PWV. Bland-Altman analysis demonstrated a non-significant bias of 0.19±0.54 m/s of ht-PWV with respect to cf-PWV. Conclusion: We conclude that, in this small-sized young and healthy population, fully automated measurement of heart-thigh PWV is straightforward and easy. Measured values were not different from carotid-femoral PWV. Further research is warranted to confirm these findings in a larger population spanning a large age range and cardiovascular risk profiles.
url https://www.atlantis-press.com/article/125930628/view
work_keys_str_mv AT franciscolondono p514thighcuffbasedmeasurementofaorticpulsewavevelocityinitialtestingofanovelvaseraprototypedevice
AT daimecampos p514thighcuffbasedmeasurementofaorticpulsewavevelocityinitialtestingofanovelvaseraprototypedevice
AT shigeohorinaka p514thighcuffbasedmeasurementofaorticpulsewavevelocityinitialtestingofanovelvaseraprototypedevice
AT juliochirinos p514thighcuffbasedmeasurementofaorticpulsewavevelocityinitialtestingofanovelvaseraprototypedevice
AT patricksegers p514thighcuffbasedmeasurementofaorticpulsewavevelocityinitialtestingofanovelvaseraprototypedevice
_version_ 1724587512692736000