Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode Switch

An 80-year-old patient having a dual chamber DR pacer connected to a VDD lead presented with chronic lead dislodgment with the atrial ring displaced into the right ventricle. There was no ventricular capture at maximal ventricular output, and given the clinical settings, the condition was managed wi...

Full description

Bibliographic Details
Main Authors: Antoine Kossaify, Nayla Nicolas
Format: Article
Language:English
Published: SAGE Publishing 2012-01-01
Series:Clinical Medicine Insights: Case Reports
Online Access:https://doi.org/10.4137/CCRep.S9625
id doaj-2fbc1b60473c4fdc84398b571ffb5bc4
record_format Article
spelling doaj-2fbc1b60473c4fdc84398b571ffb5bc42020-11-25T03:21:39ZengSAGE PublishingClinical Medicine Insights: Case Reports1179-54762012-01-01510.4137/CCRep.S9625Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode SwitchAntoine Kossaify0Nayla Nicolas1Electrophysiology Division, Cardiology department, USEK/NDS–-University Hospital, St. Charbel Street, Byblos, Lebanon.Electrophysiology Division, Cardiology department, USEK/NDS–-University Hospital, St. Charbel Street, Byblos, Lebanon.An 80-year-old patient having a dual chamber DR pacer connected to a VDD lead presented with chronic lead dislodgment with the atrial ring displaced into the right ventricle. There was no ventricular capture at maximal ventricular output, and given the clinical settings, the condition was managed with a conservative approach, the mode was switched to AAI ensuring a minimal adequate ventricular pacing backup in a non dependent patient.https://doi.org/10.4137/CCRep.S9625
collection DOAJ
language English
format Article
sources DOAJ
author Antoine Kossaify
Nayla Nicolas
spellingShingle Antoine Kossaify
Nayla Nicolas
Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode Switch
Clinical Medicine Insights: Case Reports
author_facet Antoine Kossaify
Nayla Nicolas
author_sort Antoine Kossaify
title Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode Switch
title_short Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode Switch
title_full Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode Switch
title_fullStr Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode Switch
title_full_unstemmed Dislodgment of VDD Lead Connected to DR Pacer and Managed by Mode Switch
title_sort dislodgment of vdd lead connected to dr pacer and managed by mode switch
publisher SAGE Publishing
series Clinical Medicine Insights: Case Reports
issn 1179-5476
publishDate 2012-01-01
description An 80-year-old patient having a dual chamber DR pacer connected to a VDD lead presented with chronic lead dislodgment with the atrial ring displaced into the right ventricle. There was no ventricular capture at maximal ventricular output, and given the clinical settings, the condition was managed with a conservative approach, the mode was switched to AAI ensuring a minimal adequate ventricular pacing backup in a non dependent patient.
url https://doi.org/10.4137/CCRep.S9625
work_keys_str_mv AT antoinekossaify dislodgmentofvddleadconnectedtodrpacerandmanagedbymodeswitch
AT naylanicolas dislodgmentofvddleadconnectedtodrpacerandmanagedbymodeswitch
_version_ 1724613512978759680