Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus
A portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska. We compared serum chemistry results obtained on-site with results from a tradition...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Hindawi Limited
2010-01-01
|
Series: | Veterinary Medicine International |
Online Access: | http://dx.doi.org/10.4061/2010/418596 |
id |
doaj-3df0596964cb4ee2a283fdaa12e6f60b |
---|---|
record_format |
Article |
spelling |
doaj-3df0596964cb4ee2a283fdaa12e6f60b2020-11-24T21:38:20ZengHindawi LimitedVeterinary Medicine International2042-00482010-01-01201010.4061/2010/418596418596Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicusMichael K. Stoskopf0Daniel M. Mulcahy1Daniel Esler2Environmental Medicine Consortium and Department of Clinical Sciences, College of Veterinary Medicine, North Carolina State University, 4700 Hillsborough Street, Raleigh, NC 27606, USAU.S. Geological Survey, Alaska Science Center, 4210 University Drive, Anchorage, AK 99508, USAU.S. Geological Survey, Alaska Science Center, 4210 University Drive, Anchorage, AK 99508, USAA portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska. We compared serum chemistry results obtained on-site with results from a traditional laboratory. Particular attention was paid to serum glucose and potassium concentrations as potential indicators of high-risk surgical candidates based on evaluation of the field data. The median differential for glucose values (N=82) between methods was 0.6 mmol/L (quartiles 0.3 and 0.9 mmol/L) with the median value higher when assayed on site. Analysis of potassium on site returned a median of 2.7 mmol/L (N=88 ; quartiles 2.4 and 3.0 mmol/L). Serum potassium values were too low for quantitation by the traditional laboratory. Changes in several serum chemistry values following a three-day storm during the study support the value of on site evaluation of serum potassium to identify presurgical patients with increased anesthetic risk.http://dx.doi.org/10.4061/2010/418596 |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Michael K. Stoskopf Daniel M. Mulcahy Daniel Esler |
spellingShingle |
Michael K. Stoskopf Daniel M. Mulcahy Daniel Esler Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus Veterinary Medicine International |
author_facet |
Michael K. Stoskopf Daniel M. Mulcahy Daniel Esler |
author_sort |
Michael K. Stoskopf |
title |
Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus |
title_short |
Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus |
title_full |
Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus |
title_fullStr |
Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus |
title_full_unstemmed |
Evaluation of a Portable Automated Serum Chemistry Analyzer for Field Assessment of Harlequin Ducks, Histrionicus histrionicus |
title_sort |
evaluation of a portable automated serum chemistry analyzer for field assessment of harlequin ducks, histrionicus histrionicus |
publisher |
Hindawi Limited |
series |
Veterinary Medicine International |
issn |
2042-0048 |
publishDate |
2010-01-01 |
description |
A portable analytical chemistry analyzer was used to make field assessments of wild harlequin ducks (Histrionicus histrionicus) in association with telemetry studies of winter survival in Prince William Sound, Alaska. We compared serum chemistry results obtained on-site with results from a traditional laboratory. Particular attention was paid to serum glucose and potassium concentrations as potential indicators of high-risk surgical candidates based on evaluation of the field data. The median differential for glucose values (N=82) between methods was 0.6 mmol/L (quartiles 0.3 and 0.9 mmol/L) with the median value higher when assayed on site. Analysis of potassium on site returned a median of 2.7 mmol/L (N=88
; quartiles 2.4 and 3.0 mmol/L). Serum potassium values were too low for quantitation by the traditional laboratory. Changes in several serum chemistry values following a three-day storm during the study support the value of on site evaluation of serum potassium to identify presurgical patients with increased anesthetic risk. |
url |
http://dx.doi.org/10.4061/2010/418596 |
work_keys_str_mv |
AT michaelkstoskopf evaluationofaportableautomatedserumchemistryanalyzerforfieldassessmentofharlequinduckshistrionicushistrionicus AT danielmmulcahy evaluationofaportableautomatedserumchemistryanalyzerforfieldassessmentofharlequinduckshistrionicushistrionicus AT danielesler evaluationofaportableautomatedserumchemistryanalyzerforfieldassessmentofharlequinduckshistrionicushistrionicus |
_version_ |
1725934812354576384 |