Comparison of anterior chamber depths measured using the Pentacam, the IOLMaster, and ultrasound pachymetry

AIM: To investigate and compare anterior chamber depths(ACDs)measured using the Pentacam, the IOLMaster, and ultrasound pachymetry(US).<p>METHODS: The present study was observational in nature. ACDs were measured in 138 eyes of 69 myopic patients, by the same operator, using the Pentacam, the...

Full description

Bibliographic Details
Main Authors: Jing Zhang, Shi-Sheng Zhang, Qing Yu, Jiang-Xiu Wu, Jing-Cai Lian
Format: Article
Language:English
Published: Press of International Journal of Ophthalmology (IJO PRESS) 2015-08-01
Series:Guoji Yanke Zazhi
Subjects:
Online Access:http://ies.ijo.cn/cn_publish/2015/8/201508003.pdf
id doaj-482bd3b9c140430c83bc44b46d103b5f
record_format Article
spelling doaj-482bd3b9c140430c83bc44b46d103b5f2020-11-24T22:54:29ZengPress of International Journal of Ophthalmology (IJO PRESS)Guoji Yanke Zazhi1672-51231672-51232015-08-011581313131810.3980/j.issn.1672-5123.2015.8.03Comparison of anterior chamber depths measured using the Pentacam, the IOLMaster, and ultrasound pachymetryJing Zhang0Shi-Sheng Zhang1Qing Yu2Jiang-Xiu Wu3Jing-Cai Lian4Department of Ophthalmology, Shanghai Ninth People's Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200011, ChinaNew Vision Eye Clinic, Shanghai 200011, ChinaNew Vision Eye Clinic, Shanghai 200011, ChinaNew Vision Eye Clinic, Shanghai 200011, ChinaNew Vision Eye Clinic, Shanghai 200011, China;Department of Ophthalmology, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 200025, ChinaAIM: To investigate and compare anterior chamber depths(ACDs)measured using the Pentacam, the IOLMaster, and ultrasound pachymetry(US).<p>METHODS: The present study was observational in nature. ACDs were measured in 138 eyes of 69 myopic patients, by the same operator, using the Pentacam, the IOLMaster, and US. We assessed the agreement among the three methods using Bland-Altman plots. The repeatability among the three methods was evaluated by within-subject standard deviation. <p>RESULTS: The means±SDs of ACDs measured using the Pentacam, the IOLMaster, and US were 3.77±0.24, 3.73±0.23, and 3.69±0.22mm respectively. Bland-Altman analysis showed that Pentacam and IOLMaster data were in good agreement(CoA, 0.04mm; LoA,0.05 to 0.13mm), as were US and IOLMaster data(CoA, 0.04mm; LoA, 0.17 to 0.08mm), but the Pentacam ACD values were slightly greater than the US figures(CoA, 0.08mm; LoA, 0.06 to 0.22mm). Measurements of the ACD with the three devices also showed high repeatability(S<sub>w</sub>=0.03, 0.02, and 0.03; 2.77 S<sub>w</sub>=0.08, 0.06, and 0.08, respectively). The three depth estimates were positively correlated(<i>r</i>=0.946, 0.987, and 0.951; <i>P</i><0.001).<p>CONCLUSION: Measurement of ACDs using the Pentacam, the IOLMaster, and US showed good agreement and repeatability. The Pentacam and IOLMaster ACDs, and the IOLMaster and US ACDs, agreed reasonably well, but agreement between the Pentacam and US data was poorer. However, both the absolute differences and the coefficients of variation were small, and the observed variability was likely not clinically significant.http://ies.ijo.cn/cn_publish/2015/8/201508003.pdfanterior chamber depthPentacamIOLMasterultrasound pachymetry
collection DOAJ
language English
format Article
sources DOAJ
author Jing Zhang
Shi-Sheng Zhang
Qing Yu
Jiang-Xiu Wu
Jing-Cai Lian
spellingShingle Jing Zhang
Shi-Sheng Zhang
Qing Yu
Jiang-Xiu Wu
Jing-Cai Lian
Comparison of anterior chamber depths measured using the Pentacam, the IOLMaster, and ultrasound pachymetry
Guoji Yanke Zazhi
anterior chamber depth
Pentacam
IOLMaster
ultrasound pachymetry
author_facet Jing Zhang
Shi-Sheng Zhang
Qing Yu
Jiang-Xiu Wu
Jing-Cai Lian
author_sort Jing Zhang
title Comparison of anterior chamber depths measured using the Pentacam, the IOLMaster, and ultrasound pachymetry
title_short Comparison of anterior chamber depths measured using the Pentacam, the IOLMaster, and ultrasound pachymetry
title_full Comparison of anterior chamber depths measured using the Pentacam, the IOLMaster, and ultrasound pachymetry
title_fullStr Comparison of anterior chamber depths measured using the Pentacam, the IOLMaster, and ultrasound pachymetry
title_full_unstemmed Comparison of anterior chamber depths measured using the Pentacam, the IOLMaster, and ultrasound pachymetry
title_sort comparison of anterior chamber depths measured using the pentacam, the iolmaster, and ultrasound pachymetry
publisher Press of International Journal of Ophthalmology (IJO PRESS)
series Guoji Yanke Zazhi
issn 1672-5123
1672-5123
publishDate 2015-08-01
description AIM: To investigate and compare anterior chamber depths(ACDs)measured using the Pentacam, the IOLMaster, and ultrasound pachymetry(US).<p>METHODS: The present study was observational in nature. ACDs were measured in 138 eyes of 69 myopic patients, by the same operator, using the Pentacam, the IOLMaster, and US. We assessed the agreement among the three methods using Bland-Altman plots. The repeatability among the three methods was evaluated by within-subject standard deviation. <p>RESULTS: The means±SDs of ACDs measured using the Pentacam, the IOLMaster, and US were 3.77±0.24, 3.73±0.23, and 3.69±0.22mm respectively. Bland-Altman analysis showed that Pentacam and IOLMaster data were in good agreement(CoA, 0.04mm; LoA,0.05 to 0.13mm), as were US and IOLMaster data(CoA, 0.04mm; LoA, 0.17 to 0.08mm), but the Pentacam ACD values were slightly greater than the US figures(CoA, 0.08mm; LoA, 0.06 to 0.22mm). Measurements of the ACD with the three devices also showed high repeatability(S<sub>w</sub>=0.03, 0.02, and 0.03; 2.77 S<sub>w</sub>=0.08, 0.06, and 0.08, respectively). The three depth estimates were positively correlated(<i>r</i>=0.946, 0.987, and 0.951; <i>P</i><0.001).<p>CONCLUSION: Measurement of ACDs using the Pentacam, the IOLMaster, and US showed good agreement and repeatability. The Pentacam and IOLMaster ACDs, and the IOLMaster and US ACDs, agreed reasonably well, but agreement between the Pentacam and US data was poorer. However, both the absolute differences and the coefficients of variation were small, and the observed variability was likely not clinically significant.
topic anterior chamber depth
Pentacam
IOLMaster
ultrasound pachymetry
url http://ies.ijo.cn/cn_publish/2015/8/201508003.pdf
work_keys_str_mv AT jingzhang comparisonofanteriorchamberdepthsmeasuredusingthepentacamtheiolmasterandultrasoundpachymetry
AT shishengzhang comparisonofanteriorchamberdepthsmeasuredusingthepentacamtheiolmasterandultrasoundpachymetry
AT qingyu comparisonofanteriorchamberdepthsmeasuredusingthepentacamtheiolmasterandultrasoundpachymetry
AT jiangxiuwu comparisonofanteriorchamberdepthsmeasuredusingthepentacamtheiolmasterandultrasoundpachymetry
AT jingcailian comparisonofanteriorchamberdepthsmeasuredusingthepentacamtheiolmasterandultrasoundpachymetry
_version_ 1725659565870022656