A method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatment
<p>Purpose: This study was aimed to increase the quality of high dose rate (HDR) intraluminal brachytherapy treatment. For this purpose, an easy, fast and accurate patient-specific quality assurance (QA) tool has been developed. This tool has been implemented at Bahawalpur Institute of Nuclear...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
IJCTO
2016-06-01
|
Series: | International Journal of Cancer Therapy and Oncology |
Subjects: | |
Online Access: | http://ijcto.org/index.php/IJCTO/article/view/462 |
id |
doaj-4969ea32e3a54497a77ac58376f5b2c1 |
---|---|
record_format |
Article |
spelling |
doaj-4969ea32e3a54497a77ac58376f5b2c12020-11-24T23:01:08ZengIJCTOInternational Journal of Cancer Therapy and Oncology 2330-40492016-06-014210.14319/ijcto.42.3259A method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatmentMuhammad Asghar Gadhi0Shahab Fatmi1Muhammad Shakil2Saeed Ahmad Buzdar3Department of Medical Physics, Bahawalpur Institute of Nuclear Medicine and Oncology (BINO), BahawalpurDepartment of Nuclear Medicine, Bahawalpur Institute of Nuclear Medicine and Oncology (BINO), BahawalpurDepartment of Physics, University of Gujrat, GujratMedical Physics Research Group, Department of Physics, The Islamia University of Bahawalpur, Bahawalpur<p>Purpose: This study was aimed to increase the quality of high dose rate (HDR) intraluminal brachytherapy treatment. For this purpose, an easy, fast and accurate patient-specific quality assurance (QA) tool has been developed. This tool has been implemented at Bahawalpur Institute of Nuclear Medicine and Oncology (BINO), Bahawalpur, Pakistan.</p><p>Methods: ABACUS 3.1 Treatment planning system (TPS) has been used for treatment planning and calculation of total dwell time and then results were compared with the time calculated using the proposed method. This method has been used to verify the total dwell time for different rectum applicators for relevant treatment lengths (2-7 cm) and depths (1.5-2.5 cm), different oesophagus applicators of relevant treatment lengths (6-10 cm) and depths (0.9 & 1.0 cm), and a bronchus applicator for relevant treatment lengths (4-7.5 cm) and depth (0.5 cm).</p><p>Results: The average percentage differences between treatment time T<sub>M </sub>with manual calculation and as calculated by the TPS is 0.32% (standard deviation 1.32%) for rectum, 0.24% (standard deviation 2.36%) for oesophagus and 1.96% (standard deviation 0.55%) for bronchus, respectively. These results advocate that the proposed method is valuable for independent verification of patient-specific treatment planning QA.</p><p>Conclusion: The technique illustrated in the current study is an easy, simple, quick and useful for independent verification of the total dwell time for HDR intraluminal brachytherapy. Our method is able to identify human error-related planning mistakes and to evaluate the quality of treatment planning. It enhances the quality of brachytherapy treatment and reliability of the system.</p>http://ijcto.org/index.php/IJCTO/article/view/462Intraluminal, Brachytherapy, Independent Method, High Dose Rate |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Muhammad Asghar Gadhi Shahab Fatmi Muhammad Shakil Saeed Ahmad Buzdar |
spellingShingle |
Muhammad Asghar Gadhi Shahab Fatmi Muhammad Shakil Saeed Ahmad Buzdar A method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatment International Journal of Cancer Therapy and Oncology Intraluminal, Brachytherapy, Independent Method, High Dose Rate |
author_facet |
Muhammad Asghar Gadhi Shahab Fatmi Muhammad Shakil Saeed Ahmad Buzdar |
author_sort |
Muhammad Asghar Gadhi |
title |
A method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatment |
title_short |
A method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatment |
title_full |
A method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatment |
title_fullStr |
A method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatment |
title_full_unstemmed |
A method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatment |
title_sort |
method for verification of treatment times for high-dose-rate intraluminal brachytherapy treatment |
publisher |
IJCTO |
series |
International Journal of Cancer Therapy and Oncology |
issn |
2330-4049 |
publishDate |
2016-06-01 |
description |
<p>Purpose: This study was aimed to increase the quality of high dose rate (HDR) intraluminal brachytherapy treatment. For this purpose, an easy, fast and accurate patient-specific quality assurance (QA) tool has been developed. This tool has been implemented at Bahawalpur Institute of Nuclear Medicine and Oncology (BINO), Bahawalpur, Pakistan.</p><p>Methods: ABACUS 3.1 Treatment planning system (TPS) has been used for treatment planning and calculation of total dwell time and then results were compared with the time calculated using the proposed method. This method has been used to verify the total dwell time for different rectum applicators for relevant treatment lengths (2-7 cm) and depths (1.5-2.5 cm), different oesophagus applicators of relevant treatment lengths (6-10 cm) and depths (0.9 & 1.0 cm), and a bronchus applicator for relevant treatment lengths (4-7.5 cm) and depth (0.5 cm).</p><p>Results: The average percentage differences between treatment time T<sub>M </sub>with manual calculation and as calculated by the TPS is 0.32% (standard deviation 1.32%) for rectum, 0.24% (standard deviation 2.36%) for oesophagus and 1.96% (standard deviation 0.55%) for bronchus, respectively. These results advocate that the proposed method is valuable for independent verification of patient-specific treatment planning QA.</p><p>Conclusion: The technique illustrated in the current study is an easy, simple, quick and useful for independent verification of the total dwell time for HDR intraluminal brachytherapy. Our method is able to identify human error-related planning mistakes and to evaluate the quality of treatment planning. It enhances the quality of brachytherapy treatment and reliability of the system.</p> |
topic |
Intraluminal, Brachytherapy, Independent Method, High Dose Rate |
url |
http://ijcto.org/index.php/IJCTO/article/view/462 |
work_keys_str_mv |
AT muhammadasghargadhi amethodforverificationoftreatmenttimesforhighdoserateintraluminalbrachytherapytreatment AT shahabfatmi amethodforverificationoftreatmenttimesforhighdoserateintraluminalbrachytherapytreatment AT muhammadshakil amethodforverificationoftreatmenttimesforhighdoserateintraluminalbrachytherapytreatment AT saeedahmadbuzdar amethodforverificationoftreatmenttimesforhighdoserateintraluminalbrachytherapytreatment AT muhammadasghargadhi methodforverificationoftreatmenttimesforhighdoserateintraluminalbrachytherapytreatment AT shahabfatmi methodforverificationoftreatmenttimesforhighdoserateintraluminalbrachytherapytreatment AT muhammadshakil methodforverificationoftreatmenttimesforhighdoserateintraluminalbrachytherapytreatment AT saeedahmadbuzdar methodforverificationoftreatmenttimesforhighdoserateintraluminalbrachytherapytreatment |