A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population

Introduction: Automated blood pressure (BP) monitor is widely used to assess the blood pressure (BP) of the study subjects in community-based researches. This study aims at the detection of hypertension by automated BP device and examines the concordance and validity between automated and mercury-fr...

Full description

Bibliographic Details
Main Authors: Shashi B Singh, Dewesh Kumar, Vivek Kashyap, Surendra Singh
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2020-01-01
Series:Journal of Family Medicine and Primary Care
Subjects:
Online Access:http://www.jfmpc.com/article.asp?issn=2249-4863;year=2020;volume=9;issue=3;spage=1464;epage=1469;aulast=Singh
id doaj-517c883f450b4debacf90dd8ef7e9bee
record_format Article
spelling doaj-517c883f450b4debacf90dd8ef7e9bee2020-11-25T03:37:31ZengWolters Kluwer Medknow PublicationsJournal of Family Medicine and Primary Care2249-48632020-01-01931464146910.4103/jfmpc.jfmpc_796_19A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian populationShashi B SinghDewesh KumarVivek KashyapSurendra SinghIntroduction: Automated blood pressure (BP) monitor is widely used to assess the blood pressure (BP) of the study subjects in community-based researches. This study aims at the detection of hypertension by automated BP device and examines the concordance and validity between automated and mercury-free LED BP devices. Materials and Methods: This cross-sectional study was conducted in the tribal state of Jharkhand in India from January 2017 to June 2017. A total of 300 study participants aged more than 18 years were enrolled in this study. BP of the patients in the sitting position was measured three times each by automated device and mercury-free LED BP device. The different sets of readings were assessed by concordance correlation coefficient (CCC) and other validity measures. Results: The CCC for systolic and diastolic BP measured by automated and mercury-free LED BP is 0.88 and 0.85, respectively. The mean difference between systolic and diastolic BP by both the instruments is statistically insignificant (P > 0.05). The sensitivity, specificity, PPV, NPV, and accuracy of automated BP devices to predict hypertension is 96.61%, 92.21%, 75%, 99%, and 93%. The area under ROC for systolic and diastolic BP is 0.984 and 0.97, respectively with P values < 0.0001 in both the cases. Conclusions: This study concluded that the overall automated BP machine has fair degree of agreement (CCC) with a manual BP device. The validity of this monitor to screen hypertension may also be considered in field settings.http://www.jfmpc.com/article.asp?issn=2249-4863;year=2020;volume=9;issue=3;spage=1464;epage=1469;aulast=Singhautomated bp deviceconcordance correlation coefficientkappa statisticmercury-free led bp instrumentvalidity
collection DOAJ
language English
format Article
sources DOAJ
author Shashi B Singh
Dewesh Kumar
Vivek Kashyap
Surendra Singh
spellingShingle Shashi B Singh
Dewesh Kumar
Vivek Kashyap
Surendra Singh
A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
Journal of Family Medicine and Primary Care
automated bp device
concordance correlation coefficient
kappa statistic
mercury-free led bp instrument
validity
author_facet Shashi B Singh
Dewesh Kumar
Vivek Kashyap
Surendra Singh
author_sort Shashi B Singh
title A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_short A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_full A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_fullStr A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_full_unstemmed A comparative study of automated blood pressure device and mercury-free LED blood pressure device using Lin's concordance correlation coefficient and other validity measures in Indian population
title_sort comparative study of automated blood pressure device and mercury-free led blood pressure device using lin's concordance correlation coefficient and other validity measures in indian population
publisher Wolters Kluwer Medknow Publications
series Journal of Family Medicine and Primary Care
issn 2249-4863
publishDate 2020-01-01
description Introduction: Automated blood pressure (BP) monitor is widely used to assess the blood pressure (BP) of the study subjects in community-based researches. This study aims at the detection of hypertension by automated BP device and examines the concordance and validity between automated and mercury-free LED BP devices. Materials and Methods: This cross-sectional study was conducted in the tribal state of Jharkhand in India from January 2017 to June 2017. A total of 300 study participants aged more than 18 years were enrolled in this study. BP of the patients in the sitting position was measured three times each by automated device and mercury-free LED BP device. The different sets of readings were assessed by concordance correlation coefficient (CCC) and other validity measures. Results: The CCC for systolic and diastolic BP measured by automated and mercury-free LED BP is 0.88 and 0.85, respectively. The mean difference between systolic and diastolic BP by both the instruments is statistically insignificant (P > 0.05). The sensitivity, specificity, PPV, NPV, and accuracy of automated BP devices to predict hypertension is 96.61%, 92.21%, 75%, 99%, and 93%. The area under ROC for systolic and diastolic BP is 0.984 and 0.97, respectively with P values < 0.0001 in both the cases. Conclusions: This study concluded that the overall automated BP machine has fair degree of agreement (CCC) with a manual BP device. The validity of this monitor to screen hypertension may also be considered in field settings.
topic automated bp device
concordance correlation coefficient
kappa statistic
mercury-free led bp instrument
validity
url http://www.jfmpc.com/article.asp?issn=2249-4863;year=2020;volume=9;issue=3;spage=1464;epage=1469;aulast=Singh
work_keys_str_mv AT shashibsingh acomparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT deweshkumar acomparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT vivekkashyap acomparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT surendrasingh acomparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT shashibsingh comparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT deweshkumar comparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT vivekkashyap comparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
AT surendrasingh comparativestudyofautomatedbloodpressuredeviceandmercuryfreeledbloodpressuredeviceusinglinsconcordancecorrelationcoefficientandothervaliditymeasuresinindianpopulation
_version_ 1724545485509754880