Summary: | The paper presents the influence of mechanical surface damage on the physicochemical properties of plates after implantation made of CrNiMo stainless steel, used in the treatment of anterior surface deformity of the chest. Analysis of the data allowed us to investigate the effect of implant design and condition of their surface on the results of chest deformation treatment. Results of electrochemical, impedance and surface wettability tests and SEM observations were compared with clinical observations. When removing the plates we found only slight inflammatory-periosteal reactions. On the basis of obtained results, it can be stated that plates, in spite of mechanical damage of the surface, were characterized by good corrosion resistance, a fact which is confirmed by the results of clinical evaluation.
|