Comparison of intraocular pressure measurements between Icare PRO Tonometer, Goldmann Applanation Tonometer and non-contact tonometer in healthy and glaucomatous eyes
Abstract Purpose: To validate the comparative measurements of intraocular pressure performed with IcarePRO tonometer (IPT) in relation to Goldmann applanation tonometer (GAT) and non-contact tonometer (NCT), as well as to evaluate the influence of central corneal thickness (CCT) on these values. M...
Main Authors: | Tatiana Sofia Monteiro Queirós, Márcia Clara Sá Rodrigues, Ana Carolina Pereira do Vale, Maria Josefina Costa Serino, Maria Teresa Leal Felgueiras Paínhas, Gabriel José Morgado Santos, Nuno Pedro Filipe da Costa e Lopes |
---|---|
Format: | Article |
Language: | English |
Published: |
Sociedade Brasileira de Oftalmologia
|
Series: | Revista Brasileira de Oftalmologia |
Subjects: | |
Online Access: | http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0034-72802018000500248&lng=en&tlng=en |
Similar Items
-
Intraocular pressure measurement in sheep using an applanation tonometer
by: João Antonio Tadeu Pigatto, et al.
Published: (2011-12-01) -
Comparison of the Accuracy of Intraocular Pressure Measurement by Goldmann Applanation Tonometer and Keeler Noncontact Tonometer
by: Noushin Bazzazi, et al.
Published: (2013-07-01) -
COMPARISON OF INTRAOCULAR PRESSURE MEASURED WITH GOLDMANN APPLANATION TONOMETER AND NON-CONTACT AIR-PUFF TONOMETER
by: Hidayatullah Mahsud, et al.
Published: (2016-07-01) -
Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma
by: Shunsuke Nakakura, et al.
Published: (2021-07-01) -
A comparison of intraocular pressure measurements using rebound tonometry (iCare® tonometer) and applanation tonometry (Goldmann tonometer) in a South African clinical setting
Published: (2015)