Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012

National Electrical Manufacturers Association (NEMA) provides guidelines to assess the performance of Positron Emission Tomography (PET). A PET/CT scanner, Discovery IQ, GE Medical systems, Milwaukee, USA was installed in our department which has high a sensitivity PET component. We have performed t...

Full description

Bibliographic Details
Main Authors: Ashish Kumar Jha, Sneha Mithun, Ameya D Puranik, Nilendu C Purandare, Sneha Shah, Archi Agrawal, Venkatesh Rangarajan
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2019-01-01
Series:World Journal of Nuclear Medicine
Subjects:
Online Access:http://www.wjnm.org/article.asp?issn=1450-1147;year=2019;volume=18;issue=4;spage=351;epage=360;aulast=Jha
id doaj-5fbe654543f546acaaa9f0fe80f6a95c
record_format Article
spelling doaj-5fbe654543f546acaaa9f0fe80f6a95c2020-11-25T01:57:46ZengWolters Kluwer Medknow PublicationsWorld Journal of Nuclear Medicine1450-11471607-33122019-01-0118435136010.4103/wjnm.WJNM_72_18Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012Ashish Kumar JhaSneha MithunAmeya D PuranikNilendu C PurandareSneha ShahArchi AgrawalVenkatesh RangarajanNational Electrical Manufacturers Association (NEMA) provides guidelines to assess the performance of Positron Emission Tomography (PET). A PET/CT scanner, Discovery IQ, GE Medical systems, Milwaukee, USA was installed in our department which has high a sensitivity PET component. We have performed the NEMA NU-2 2012 quality control tests to evaluate this system on site before clinical use. Performance measurements of the PET scanner were made using the NEMA NU2-2012 procedures for spatial resolution, scatter fraction, sensitivity, count rate loss and random coincidence estimation, Noise Equivalent Count Rate (NECR) and image quality. As per NU2 2012, spatial resolution was measured at 1 cm, 10 cm and 20 cm vertically from the centre and at each of these points resolution was measured at tangential, radial and axial directions. Sensitivity was measured at centre and 10 cm off center vertically from the center. The system sensitivity is reported as an average of the two measured values. Scatter fraction and NECR measurements, Image quality test was also performed. The tangential, radial and axial FWHM were 4.99 mm, 4.20 mm and 4.79 mm at 1 cm off centre, 5.49 mm, 4.69 mm and 4.81 mm at 10 cm off centre and 7.99 mm, 5.07 mm and 4.95 mm at 20 cm off centre respectively. The absolute sensitivity of this scanner was found to be 20.1 cps/kBq. The scatter fraction calculated from the decay method was 37.94% and NECR was 125 kcps. The peak NECR was achieved at activity concentration of 8.7 KBq/ml and the count loss below the peak NECR was found to be 0.68%. Image quality test for, contrast recovery, background variability and lung error residual mean met all specifications. Overall PET performance of Discovery IQ whole-body scanner was satisfactory and the scanner met all the performance specifications required by NEMA 2012.http://www.wjnm.org/article.asp?issn=1450-1147;year=2019;volume=18;issue=4;spage=351;epage=360;aulast=Jhadiscovery image qualityimage qualitynational electrical manufacturers association evaluationnational electrical manufacturers association nu-2 2012q. clearvphd
collection DOAJ
language English
format Article
sources DOAJ
author Ashish Kumar Jha
Sneha Mithun
Ameya D Puranik
Nilendu C Purandare
Sneha Shah
Archi Agrawal
Venkatesh Rangarajan
spellingShingle Ashish Kumar Jha
Sneha Mithun
Ameya D Puranik
Nilendu C Purandare
Sneha Shah
Archi Agrawal
Venkatesh Rangarajan
Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
World Journal of Nuclear Medicine
discovery image quality
image quality
national electrical manufacturers association evaluation
national electrical manufacturers association nu-2 2012
q. clear
vphd
author_facet Ashish Kumar Jha
Sneha Mithun
Ameya D Puranik
Nilendu C Purandare
Sneha Shah
Archi Agrawal
Venkatesh Rangarajan
author_sort Ashish Kumar Jha
title Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_short Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_full Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_fullStr Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_full_unstemmed Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_sort performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per national electrical manufacturers association nu 2-2012
publisher Wolters Kluwer Medknow Publications
series World Journal of Nuclear Medicine
issn 1450-1147
1607-3312
publishDate 2019-01-01
description National Electrical Manufacturers Association (NEMA) provides guidelines to assess the performance of Positron Emission Tomography (PET). A PET/CT scanner, Discovery IQ, GE Medical systems, Milwaukee, USA was installed in our department which has high a sensitivity PET component. We have performed the NEMA NU-2 2012 quality control tests to evaluate this system on site before clinical use. Performance measurements of the PET scanner were made using the NEMA NU2-2012 procedures for spatial resolution, scatter fraction, sensitivity, count rate loss and random coincidence estimation, Noise Equivalent Count Rate (NECR) and image quality. As per NU2 2012, spatial resolution was measured at 1 cm, 10 cm and 20 cm vertically from the centre and at each of these points resolution was measured at tangential, radial and axial directions. Sensitivity was measured at centre and 10 cm off center vertically from the center. The system sensitivity is reported as an average of the two measured values. Scatter fraction and NECR measurements, Image quality test was also performed. The tangential, radial and axial FWHM were 4.99 mm, 4.20 mm and 4.79 mm at 1 cm off centre, 5.49 mm, 4.69 mm and 4.81 mm at 10 cm off centre and 7.99 mm, 5.07 mm and 4.95 mm at 20 cm off centre respectively. The absolute sensitivity of this scanner was found to be 20.1 cps/kBq. The scatter fraction calculated from the decay method was 37.94% and NECR was 125 kcps. The peak NECR was achieved at activity concentration of 8.7 KBq/ml and the count loss below the peak NECR was found to be 0.68%. Image quality test for, contrast recovery, background variability and lung error residual mean met all specifications. Overall PET performance of Discovery IQ whole-body scanner was satisfactory and the scanner met all the performance specifications required by NEMA 2012.
topic discovery image quality
image quality
national electrical manufacturers association evaluation
national electrical manufacturers association nu-2 2012
q. clear
vphd
url http://www.wjnm.org/article.asp?issn=1450-1147;year=2019;volume=18;issue=4;spage=351;epage=360;aulast=Jha
work_keys_str_mv AT ashishkumarjha performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT snehamithun performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT ameyadpuranik performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT nilenducpurandare performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT snehashah performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT archiagrawal performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT venkateshrangarajan performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
_version_ 1724972646648512512