Development of a Novel, Low-Cost, Disposable Wooden Pencil Graphite Electrode for Use in the Determination of Antioxidants and Other Biological Compounds

The development of portable sensors that can be used outside the lab is an active area of research in the electroanalytical field. A major focus of such research is the development of low-cost electrodes for use in these sensors. Current electrodes, such as glassy-carbon electrodes (GCEs), are costl...

Full description

Bibliographic Details
Main Authors: James Kariuki, Emily Ervin, Carly Olafson
Format: Article
Language:English
Published: MDPI AG 2015-07-01
Series:Sensors
Subjects:
Online Access:http://www.mdpi.com/1424-8220/15/8/18887
Description
Summary:The development of portable sensors that can be used outside the lab is an active area of research in the electroanalytical field. A major focus of such research is the development of low-cost electrodes for use in these sensors. Current electrodes, such as glassy-carbon electrodes (GCEs), are costly and require time-consuming preparation. Alternatives have been proposed, including mechanical pencil-lead electrodes (MPEs). However, MPEs themselves possess numerous drawbacks, particularly structural fragility. In this paper, we present a novel pencil-graphite electrode (PGE) fabricated from a regular HB#2 pencil. This PGE is a simple, disposable, extremely low-cost alternative to GCEs ($0.30 per PGE, vs. $190 + per GCE), and possesses the structural stability that MPEs lack. PGEs were characterized by square-wave voltammetry of ferricyanide, gallic acid, uric acid, dopamine, and several foodstuffs. In all cases, PGEs demonstrated sensitivities comparable or superior to those of the GCE and MPE (LOD = 5.62 × 10−4 M PGE, 4.80 × 10−4 M GCE, 2.93 × 10−4 M MPE). Signal areas and peak heights were typically four to ten times larger for the PGE relative to the GCE.
ISSN:1424-8220