Evaluation of Large-Area Silicon Photomultiplier Arrays for Positron Emission Tomography Systems

An individual readout of silicon photomultipliers (SiPMs) would enhance the performance of modern positron emission tomography (PET) systems. However, as it difficult to achieve in practice, a multiplexing readout of SiPM arrays could be performed instead. In this study, we characterized the perform...

Full description

Bibliographic Details
Main Authors: Minjee Seo, Haewook Park, Jae Sung Lee
Format: Article
Language:English
Published: MDPI AG 2021-03-01
Series:Electronics
Subjects:
Online Access:https://www.mdpi.com/2079-9292/10/6/698
Description
Summary:An individual readout of silicon photomultipliers (SiPMs) would enhance the performance of modern positron emission tomography (PET) systems. However, as it difficult to achieve in practice, a multiplexing readout of SiPM arrays could be performed instead. In this study, we characterized the performance of three PET detector modules utilizing three different SiPM models with active areas of 3 × 3, 4 × 4, and 6 × 6 mm<sup>2</sup>. Each SiPM array was coupled with a 4 × 4 LYSO crystal block. For SiPM multiplexing, we used a discretized positioning circuit to obtain position and energy information, and applied a first-order capacitive high-pass filter to enhance the time-of-flight measurement capability of the PET detector. The energy performance was similar among the three different SiPM arrays, with an energy resolution of 10%–11%. The best timing performance was achieved with the SiPM array with an active area of 6 × 6 mm<sup>2</sup>, which yielded a coincidence timing resolution (CTR) value of 401 ps FWHM when an analog high-pass filter was applied. We expect that, in combination with high-performance SiPM multiplexing techniques, the SiPM array with an active area of 6 × 6 mm<sup>2</sup> can provide a cost-effective solution for developing a whole-body PET scanner.
ISSN:2079-9292