Study of the Microfocus X-Ray Tube Based on a Point-Like Target Used for Micro-Computed Tomography.

For a micro-Computed Tomography (Micro-CT) system, the microfocus X-ray tube is an essential component because the spatial resolution of CT images, in theory, is mainly determined by the size and stability of the X-ray focal spot of the microfocus X-ray tube. However, many factors, including voltage...

Full description

Bibliographic Details
Main Authors: Rifeng Zhou, Xiaojian Zhou, Xiaobin Li, Yufang Cai, Fenglin Liu
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2016-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC4889132?pdf=render
Description
Summary:For a micro-Computed Tomography (Micro-CT) system, the microfocus X-ray tube is an essential component because the spatial resolution of CT images, in theory, is mainly determined by the size and stability of the X-ray focal spot of the microfocus X-ray tube. However, many factors, including voltage fluctuations, mechanical vibrations, and temperature changes, can cause the size and the stability of the X-ray focal spot to degrade. A new microfocus X-ray tube based on a point-like micro-target in which the X-ray target is irradiated with an unfocused electron beam was investigated. EGS4 Monte Carlo simulation code was employed for the calculation of the X-ray intensity produced from the point-like micro-target and the substrate. The effects of several arrangements of the target material, target and beam size were studied. The simulation results demonstrated that if the intensity of X-rays generated at the point-like target is greater than half of the X-ray intensity produced on the substrate, the X-ray focal spot is determined in part by the point-like target rather than by the electron beam in the conventional X-ray tube. In theory, since it is able to reduce those unfavorable effects such as the electron beam trajectory swinging and the beam size changing for the microfocus X-ray tube, it could alleviate CT image artifacts caused by the X-ray focal spot shift and size change.
ISSN:1932-6203