Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkov method: a pilot study

The aim. To evaluate the reliability of blood pressure (BP) measurement results using a cuffless blood pressure device (CardioQVARK®) in comparison with the values obtained using the Korotkov method.Materials and methods. An observational cross-sectional study of 50 patients (25 men, mean age 60 ± 1...

Full description

Bibliographic Details
Main Authors: N. A. Gogiberidze, Z. N. Sagirova, N. O. Kuznetsova, D. G. Gognieva, P. Sh. Chomakhidze, H. Saner, P. Yu. Kopylov
Format: Article
Language:Russian
Published: Federal State Autonomous Educational Institution of Higher Education I.M. Sechenov First Moscow State Medical University of the Ministry of Health of the Russian Federation (Sechenov University) 2021-09-01
Series:Сеченовский вестник
Subjects:
Online Access:https://www.sechenovmedj.com/jour/article/view/212
id doaj-6a4fac2150ca49a8979280e2da4942c0
record_format Article
spelling doaj-6a4fac2150ca49a8979280e2da4942c02021-09-16T17:42:58ZrusFederal State Autonomous Educational Institution of Higher Education I.M. Sechenov First Moscow State Medical University of the Ministry of Health of the Russian Federation (Sechenov University)Сеченовский вестник2218-73322658-33482021-09-01121394910.47093/2218-7332.2021.12.1.39-49173Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkov method: a pilot studyN. A. Gogiberidze0Z. N. Sagirova1N. O. Kuznetsova2D. G. Gognieva3P. Sh. Chomakhidze4H. Saner5P. Yu. Kopylov6Sechenov First Moscow State Medical University (Sechenov University)Sechenov First Moscow State Medical University (Sechenov University)Sechenov First Moscow State Medical University (Sechenov University)Sechenov First Moscow State Medical University (Sechenov University)Sechenov First Moscow State Medical University (Sechenov University)ARTORG Center for Biomedical Engineering Research, University of Bern; University Clinic for Cardiology, University Hospital InselspitalSechenov First Moscow State Medical University (Sechenov University)The aim. To evaluate the reliability of blood pressure (BP) measurement results using a cuffless blood pressure device (CardioQVARK®) in comparison with the values obtained using the Korotkov method.Materials and methods. An observational cross-sectional study of 50 patients (25 men, mean age 60 ± 14 years) with arterial hypertension was performed. Blood pressure was measured by the Korotkov method as a standard method, and a CardioQVARK® device, made in the form of a smartphone case, was used as a new method. The device records the electrocardiogram and the photoplethysmogram. Based on the parameters of the electrocardiogram and the photoplethysmogram the systolic and diastolic blood pressure (SBP and DBP) is calculated. Correlation analysis, Student’s t-test, Bland-Altman method were used for comparing the two methods, the standard deviation of the difference and a 95% confidence interval (95% CI) were calculated.Results. There were no statistically significant differences in the mean values of SBP and DBP for the two methods. There was a strong direct relationship between SBP (r = 0.976, p < 0.0001) and DBP (r = 0.817, p < 0.0001), measured by two methods. Bias for SBP and DBP measured by the new method was: –0.5 mm Hg (95% CI: –1.7; 0.7) and –0.3 mmHg (95% CI: –1.4; 0.7), respectively. The difference in DBP measurements depended on the blood pressure level (r = 0.302, p = 0.03). The underestimation of DBP values was more pronounced for low blood pressure from 55 to 75 mm Hg. At the time of the study, 13 (26%) patients had an increase in blood pressure. The sensitivity of the new method in detecting arterial hypertension was 77% (95% CI: 46; 95), specificity 100% (95% CI: 91; 100), accuracy 94% (95% CI: 83; 99).Conclusion. The blood pressure measurement method based on the analysis of the electrocardiogram and photoplethysmogram showed reliable blood pressure measurement results in comparison with the Korotkov method.https://www.sechenovmedj.com/jour/article/view/212photoplethysmographypulse wave analysisblood pressureblood pressure measurementportable ecg and ppg monitorsmartphonecardioqvarktelemedicine
collection DOAJ
language Russian
format Article
sources DOAJ
author N. A. Gogiberidze
Z. N. Sagirova
N. O. Kuznetsova
D. G. Gognieva
P. Sh. Chomakhidze
H. Saner
P. Yu. Kopylov
spellingShingle N. A. Gogiberidze
Z. N. Sagirova
N. O. Kuznetsova
D. G. Gognieva
P. Sh. Chomakhidze
H. Saner
P. Yu. Kopylov
Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkov method: a pilot study
Сеченовский вестник
photoplethysmography
pulse wave analysis
blood pressure
blood pressure measurement
portable ecg and ppg monitor
smartphone
cardioqvark
telemedicine
author_facet N. A. Gogiberidze
Z. N. Sagirova
N. O. Kuznetsova
D. G. Gognieva
P. Sh. Chomakhidze
H. Saner
P. Yu. Kopylov
author_sort N. A. Gogiberidze
title Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkov method: a pilot study
title_short Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkov method: a pilot study
title_full Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkov method: a pilot study
title_fullStr Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkov method: a pilot study
title_full_unstemmed Comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the Korotkov method: a pilot study
title_sort comparison of cuffless blood pressure measurement using an electrocardiogram monitor with photoplethysmography function with measurement by the korotkov method: a pilot study
publisher Federal State Autonomous Educational Institution of Higher Education I.M. Sechenov First Moscow State Medical University of the Ministry of Health of the Russian Federation (Sechenov University)
series Сеченовский вестник
issn 2218-7332
2658-3348
publishDate 2021-09-01
description The aim. To evaluate the reliability of blood pressure (BP) measurement results using a cuffless blood pressure device (CardioQVARK®) in comparison with the values obtained using the Korotkov method.Materials and methods. An observational cross-sectional study of 50 patients (25 men, mean age 60 ± 14 years) with arterial hypertension was performed. Blood pressure was measured by the Korotkov method as a standard method, and a CardioQVARK® device, made in the form of a smartphone case, was used as a new method. The device records the electrocardiogram and the photoplethysmogram. Based on the parameters of the electrocardiogram and the photoplethysmogram the systolic and diastolic blood pressure (SBP and DBP) is calculated. Correlation analysis, Student’s t-test, Bland-Altman method were used for comparing the two methods, the standard deviation of the difference and a 95% confidence interval (95% CI) were calculated.Results. There were no statistically significant differences in the mean values of SBP and DBP for the two methods. There was a strong direct relationship between SBP (r = 0.976, p < 0.0001) and DBP (r = 0.817, p < 0.0001), measured by two methods. Bias for SBP and DBP measured by the new method was: –0.5 mm Hg (95% CI: –1.7; 0.7) and –0.3 mmHg (95% CI: –1.4; 0.7), respectively. The difference in DBP measurements depended on the blood pressure level (r = 0.302, p = 0.03). The underestimation of DBP values was more pronounced for low blood pressure from 55 to 75 mm Hg. At the time of the study, 13 (26%) patients had an increase in blood pressure. The sensitivity of the new method in detecting arterial hypertension was 77% (95% CI: 46; 95), specificity 100% (95% CI: 91; 100), accuracy 94% (95% CI: 83; 99).Conclusion. The blood pressure measurement method based on the analysis of the electrocardiogram and photoplethysmogram showed reliable blood pressure measurement results in comparison with the Korotkov method.
topic photoplethysmography
pulse wave analysis
blood pressure
blood pressure measurement
portable ecg and ppg monitor
smartphone
cardioqvark
telemedicine
url https://www.sechenovmedj.com/jour/article/view/212
work_keys_str_mv AT nagogiberidze comparisonofcufflessbloodpressuremeasurementusinganelectrocardiogrammonitorwithphotoplethysmographyfunctionwithmeasurementbythekorotkovmethodapilotstudy
AT znsagirova comparisonofcufflessbloodpressuremeasurementusinganelectrocardiogrammonitorwithphotoplethysmographyfunctionwithmeasurementbythekorotkovmethodapilotstudy
AT nokuznetsova comparisonofcufflessbloodpressuremeasurementusinganelectrocardiogrammonitorwithphotoplethysmographyfunctionwithmeasurementbythekorotkovmethodapilotstudy
AT dggognieva comparisonofcufflessbloodpressuremeasurementusinganelectrocardiogrammonitorwithphotoplethysmographyfunctionwithmeasurementbythekorotkovmethodapilotstudy
AT pshchomakhidze comparisonofcufflessbloodpressuremeasurementusinganelectrocardiogrammonitorwithphotoplethysmographyfunctionwithmeasurementbythekorotkovmethodapilotstudy
AT hsaner comparisonofcufflessbloodpressuremeasurementusinganelectrocardiogrammonitorwithphotoplethysmographyfunctionwithmeasurementbythekorotkovmethodapilotstudy
AT pyukopylov comparisonofcufflessbloodpressuremeasurementusinganelectrocardiogrammonitorwithphotoplethysmographyfunctionwithmeasurementbythekorotkovmethodapilotstudy
_version_ 1717378075884257280