Comparison of Impedance Platelet Count by Sysmex XE-5000 and Beckman Coulter LH 780 with Optical Fluorescent Platelet Count in Thalassemia Patients

Background: Red blood cell (RBC) abnormalities are frequently found among thalassemia patients and have been reported to be a cause of spuriously high platelet counts. Objective: To compare impedance platelet count by Sysmex XE-5000 (PLT-XE) and Beckman Coulter LH 780 (PLT-LH) with optical fluore...

Full description

Bibliographic Details
Main Authors: Chaicharoen Tantanate, M.D., Noppadol Siritanaratkul, M.D.
Format: Article
Language:English
Published: Mahidol University 2017-09-01
Series:Siriraj Medical Journal
Subjects:
Online Access:http://www.smj.si.mahidol.ac.th/sirirajmedj/index.php/smj/article/view/775/771
id doaj-6b26ea0ff0f949a9bc0219e2b867a015
record_format Article
spelling doaj-6b26ea0ff0f949a9bc0219e2b867a0152020-11-24T21:53:40ZengMahidol UniversitySiriraj Medical Journal2228-80822017-09-0169527628210.14456/smj.2017.54Comparison of Impedance Platelet Count by Sysmex XE-5000 and Beckman Coulter LH 780 with Optical Fluorescent Platelet Count in Thalassemia PatientsChaicharoen Tantanate, M.D.0Noppadol Siritanaratkul, M.D.1Department of Clinical Pathology, Faculty of Medicine Siriraj Hospital, Mahidol UniversityDepartment of Medicine, Faculty of Medicine Siriraj Hospital, Mahidol UniversityBackground: Red blood cell (RBC) abnormalities are frequently found among thalassemia patients and have been reported to be a cause of spuriously high platelet counts. Objective: To compare impedance platelet count by Sysmex XE-5000 (PLT-XE) and Beckman Coulter LH 780 (PLT-LH) with optical fluorescent platelet count (PLT-O) in thalassemia patients. Methods: EDTA blood specimens from 152 thalassemia patients were analyzed by routine mode of PLT-XE and PLT-LH. The PLT-O run on reticulocyte mode of Sysmex XE-5000 was used as a reference method. Spearman’s rank correlation coefficient (rs ) and Bland-Altman plot were analyzed. Results: Correlations between impedance and PLT-O methods were excellent, as demonstrated by rs of 0.95 and 0.97 for PLT-XE and PLT-LH, respectively. However, there was no agreement between impedance and PLT-O methods. There were differences in biases of PLT-XE between thalassemia minor and both thalassemia intermedia and major. Compared to PLT-O, the concordance rates of diagnosis for platelet count abnormality were 94.7% and 96.7% for PLT-XE and PLT-LH, respectively. Conclusion: Compared to PLT-O, impedance method was associated with spuriously high platelet counts in thalassemia patients. Careful blood smear examination and/or accurate assay should be used in cases suspected of spuriously high platelet count.http://www.smj.si.mahidol.ac.th/sirirajmedj/index.php/smj/article/view/775/771Hematology analyzer; impedance method; optical method; platelet count; thalassemia (Siriraj Med J 2017;69: 276-282)
collection DOAJ
language English
format Article
sources DOAJ
author Chaicharoen Tantanate, M.D.
Noppadol Siritanaratkul, M.D.
spellingShingle Chaicharoen Tantanate, M.D.
Noppadol Siritanaratkul, M.D.
Comparison of Impedance Platelet Count by Sysmex XE-5000 and Beckman Coulter LH 780 with Optical Fluorescent Platelet Count in Thalassemia Patients
Siriraj Medical Journal
Hematology analyzer; impedance method; optical method; platelet count; thalassemia (Siriraj Med J 2017;69: 276-282)
author_facet Chaicharoen Tantanate, M.D.
Noppadol Siritanaratkul, M.D.
author_sort Chaicharoen Tantanate, M.D.
title Comparison of Impedance Platelet Count by Sysmex XE-5000 and Beckman Coulter LH 780 with Optical Fluorescent Platelet Count in Thalassemia Patients
title_short Comparison of Impedance Platelet Count by Sysmex XE-5000 and Beckman Coulter LH 780 with Optical Fluorescent Platelet Count in Thalassemia Patients
title_full Comparison of Impedance Platelet Count by Sysmex XE-5000 and Beckman Coulter LH 780 with Optical Fluorescent Platelet Count in Thalassemia Patients
title_fullStr Comparison of Impedance Platelet Count by Sysmex XE-5000 and Beckman Coulter LH 780 with Optical Fluorescent Platelet Count in Thalassemia Patients
title_full_unstemmed Comparison of Impedance Platelet Count by Sysmex XE-5000 and Beckman Coulter LH 780 with Optical Fluorescent Platelet Count in Thalassemia Patients
title_sort comparison of impedance platelet count by sysmex xe-5000 and beckman coulter lh 780 with optical fluorescent platelet count in thalassemia patients
publisher Mahidol University
series Siriraj Medical Journal
issn 2228-8082
publishDate 2017-09-01
description Background: Red blood cell (RBC) abnormalities are frequently found among thalassemia patients and have been reported to be a cause of spuriously high platelet counts. Objective: To compare impedance platelet count by Sysmex XE-5000 (PLT-XE) and Beckman Coulter LH 780 (PLT-LH) with optical fluorescent platelet count (PLT-O) in thalassemia patients. Methods: EDTA blood specimens from 152 thalassemia patients were analyzed by routine mode of PLT-XE and PLT-LH. The PLT-O run on reticulocyte mode of Sysmex XE-5000 was used as a reference method. Spearman’s rank correlation coefficient (rs ) and Bland-Altman plot were analyzed. Results: Correlations between impedance and PLT-O methods were excellent, as demonstrated by rs of 0.95 and 0.97 for PLT-XE and PLT-LH, respectively. However, there was no agreement between impedance and PLT-O methods. There were differences in biases of PLT-XE between thalassemia minor and both thalassemia intermedia and major. Compared to PLT-O, the concordance rates of diagnosis for platelet count abnormality were 94.7% and 96.7% for PLT-XE and PLT-LH, respectively. Conclusion: Compared to PLT-O, impedance method was associated with spuriously high platelet counts in thalassemia patients. Careful blood smear examination and/or accurate assay should be used in cases suspected of spuriously high platelet count.
topic Hematology analyzer; impedance method; optical method; platelet count; thalassemia (Siriraj Med J 2017;69: 276-282)
url http://www.smj.si.mahidol.ac.th/sirirajmedj/index.php/smj/article/view/775/771
work_keys_str_mv AT chaicharoentantanatemd comparisonofimpedanceplateletcountbysysmexxe5000andbeckmancoulterlh780withopticalfluorescentplateletcountinthalassemiapatients
AT noppadolsiritanaratkulmd comparisonofimpedanceplateletcountbysysmexxe5000andbeckmancoulterlh780withopticalfluorescentplateletcountinthalassemiapatients
_version_ 1725870794687381504