Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography

<strong><em>Introduction:</em></strong> Digital radiography possesses a wide dynamic range and has a major advantage in producing an acceptable image of diagnostic value even though overexposure occurs. Lumbar spine (LS) radiography is the most common examinations that gives...

Full description

Bibliographic Details
Main Authors: Hanis Aisyah Ramli, Soo-Foon Moey
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2019-09-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:http://ijmp.mums.ac.ir/article_11841_3306b92c18a28a3f51924704d1a732ee.pdf
id doaj-74cc239521fe41139c1ffba41bb239fe
record_format Article
spelling doaj-74cc239521fe41139c1ffba41bb239fe2020-11-24T21:40:13ZengMashhad University of Medical SciencesIranian Journal of Medical Physics2345-36722345-36722019-09-0116536837110.22038/ijmp.2018.34167.143111841Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital RadiographyHanis Aisyah Ramli0Soo-Foon Moey1Department of Diagnostic Imaging and Radiotherapy, Kulliyyah of Allied Health Sciences, International Islamic University Malaysia, Kuantan, PahangInternational Islamic University Malaysia<strong><em>Introduction:</em></strong> Digital radiography possesses a wide dynamic range and has a major advantage in producing an acceptable image of diagnostic value even though overexposure occurs. Lumbar spine (LS) radiography is the most common examinations that gives high radiation dose to patients and accounts for the highest collective population dose of any conventional radiographic examinations. As such, this study was carried out to ascertain the impact of image quality and entrance surface dose (ESD) with different exposure settings in the anteroposterior (AP) and lateral LS. <strong><em>Material and Methods:</em></strong> The torso of the PBU-50 phantom was exposed to medium and high kilovoltage peak (kVp). A total of 14 images for LS were obtained. Relative image quality was assessed using Leeds Test Objects TOR CDR whilst the ESD was ascertained using an optically stimulated luminescence dosimeter. <strong><em>Results:</em></strong> The results of Friedman test indicated a significant difference in image quality when using medium and high kVp. Wilcoxon signed-rank test also reflected a significant difference in ESD between the use of medium and high kVp for both AP and lateral LS. <strong><em>Conclusion: </em></strong>Significant differences in image quality and ESD were obtained using medium and high kVp with medium kVp resulting in high contrast but low contrast sensitivity and vice versa. The findings of the present study indicated that the recommended kVp for AP LS was from 75kVp to 81kVp whilst for lateral LS the recommended kVp was from 85kVp to 90kVp for an average adult patient.http://ijmp.mums.ac.ir/article_11841_3306b92c18a28a3f51924704d1a732ee.pdfDigital RadiographyImage QualityRadiation Dosagelumbar spine
collection DOAJ
language English
format Article
sources DOAJ
author Hanis Aisyah Ramli
Soo-Foon Moey
spellingShingle Hanis Aisyah Ramli
Soo-Foon Moey
Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography
Iranian Journal of Medical Physics
Digital Radiography
Image Quality
Radiation Dosage
lumbar spine
author_facet Hanis Aisyah Ramli
Soo-Foon Moey
author_sort Hanis Aisyah Ramli
title Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography
title_short Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography
title_full Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography
title_fullStr Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography
title_full_unstemmed Effects of Kilovoltage on Image Quality and Entrance Surface Dose in Lumbar Spine Digital Radiography
title_sort effects of kilovoltage on image quality and entrance surface dose in lumbar spine digital radiography
publisher Mashhad University of Medical Sciences
series Iranian Journal of Medical Physics
issn 2345-3672
2345-3672
publishDate 2019-09-01
description <strong><em>Introduction:</em></strong> Digital radiography possesses a wide dynamic range and has a major advantage in producing an acceptable image of diagnostic value even though overexposure occurs. Lumbar spine (LS) radiography is the most common examinations that gives high radiation dose to patients and accounts for the highest collective population dose of any conventional radiographic examinations. As such, this study was carried out to ascertain the impact of image quality and entrance surface dose (ESD) with different exposure settings in the anteroposterior (AP) and lateral LS. <strong><em>Material and Methods:</em></strong> The torso of the PBU-50 phantom was exposed to medium and high kilovoltage peak (kVp). A total of 14 images for LS were obtained. Relative image quality was assessed using Leeds Test Objects TOR CDR whilst the ESD was ascertained using an optically stimulated luminescence dosimeter. <strong><em>Results:</em></strong> The results of Friedman test indicated a significant difference in image quality when using medium and high kVp. Wilcoxon signed-rank test also reflected a significant difference in ESD between the use of medium and high kVp for both AP and lateral LS. <strong><em>Conclusion: </em></strong>Significant differences in image quality and ESD were obtained using medium and high kVp with medium kVp resulting in high contrast but low contrast sensitivity and vice versa. The findings of the present study indicated that the recommended kVp for AP LS was from 75kVp to 81kVp whilst for lateral LS the recommended kVp was from 85kVp to 90kVp for an average adult patient.
topic Digital Radiography
Image Quality
Radiation Dosage
lumbar spine
url http://ijmp.mums.ac.ir/article_11841_3306b92c18a28a3f51924704d1a732ee.pdf
work_keys_str_mv AT hanisaisyahramli effectsofkilovoltageonimagequalityandentrancesurfacedoseinlumbarspinedigitalradiography
AT soofoonmoey effectsofkilovoltageonimagequalityandentrancesurfacedoseinlumbarspinedigitalradiography
_version_ 1725927214101299200