Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis

Abstract Background To compare the difference in central corneal thickness (CCT) measurements in normal eyes between a rotating Scheimpflug camera combined with a Placido-disk corneal topographer (Sirius, CSO, Italy) and ultrasound pachymetry (USP). Methods A systematic literature search was conduct...

Full description

Bibliographic Details
Main Authors: Yili Jin, Colm McAlinden, Yong Sun, Daizong Wen, Yiran Wang, Jinjin Yu, Ke Feng, Benhao Song, Qinmei Wang, Shihao Chen, Jinhai Huang
Format: Article
Language:English
Published: BMC 2021-02-01
Series:Eye and Vision
Subjects:
Online Access:https://doi.org/10.1186/s40662-021-00227-5
id doaj-74d3a504d8174cc4875632df19c10a5e
record_format Article
spelling doaj-74d3a504d8174cc4875632df19c10a5e2021-02-21T12:25:13ZengBMCEye and Vision2326-02542021-02-018111110.1186/s40662-021-00227-5Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysisYili Jin0Colm McAlinden1Yong Sun2Daizong Wen3Yiran Wang4Jinjin Yu5Ke Feng6Benhao Song7Qinmei Wang8Shihao Chen9Jinhai Huang10School of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversitySchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversityShenzhen Hospital of Integrated Traditional and Western MedicineSchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversitySchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversitySchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversitySchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversitySchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversitySchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversitySchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversitySchool of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical UniversityAbstract Background To compare the difference in central corneal thickness (CCT) measurements in normal eyes between a rotating Scheimpflug camera combined with a Placido-disk corneal topographer (Sirius, CSO, Italy) and ultrasound pachymetry (USP). Methods A systematic literature search was conducted for relevant studies published on PubMed, Medline, EMBASE, and the Cochrane Library and ClinicalTrials.gov from inception to August 1st, 2019. Primary outcome measures were CCT measurements between Sirius and USP. A random effects model was used to pool CCT measurements. Results A total of twelve studies involving 862 eyes were included in this meta-analysis. The meta-analysis found CCT measurements between Sirius and USP to be statistically significantly different (P < 0.0001). The mean difference between Sirius and USP was −11.26 μm with a 95% confidence interval (CI) (−16.92 μm, −5.60 μm). The heterogeneity was I2 = 60% (P = 0.004). Conclusion CCT measurements with the Sirius Scheimpflug-Placido topographer were statistically significantly lower than USP. However, it may be argued that the mean difference of 11.26 μm is not a clinically significant difference.https://doi.org/10.1186/s40662-021-00227-5Meta-analysisCentral corneal thicknessScheimpflug-Placido topographerUltrasound Pachymetry
collection DOAJ
language English
format Article
sources DOAJ
author Yili Jin
Colm McAlinden
Yong Sun
Daizong Wen
Yiran Wang
Jinjin Yu
Ke Feng
Benhao Song
Qinmei Wang
Shihao Chen
Jinhai Huang
spellingShingle Yili Jin
Colm McAlinden
Yong Sun
Daizong Wen
Yiran Wang
Jinjin Yu
Ke Feng
Benhao Song
Qinmei Wang
Shihao Chen
Jinhai Huang
Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
Eye and Vision
Meta-analysis
Central corneal thickness
Scheimpflug-Placido topographer
Ultrasound Pachymetry
author_facet Yili Jin
Colm McAlinden
Yong Sun
Daizong Wen
Yiran Wang
Jinjin Yu
Ke Feng
Benhao Song
Qinmei Wang
Shihao Chen
Jinhai Huang
author_sort Yili Jin
title Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
title_short Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
title_full Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
title_fullStr Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
title_full_unstemmed Sirius Scheimpflug–Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
title_sort sirius scheimpflug–placido versus ultrasound pachymetry for central corneal thickness: meta-analysis
publisher BMC
series Eye and Vision
issn 2326-0254
publishDate 2021-02-01
description Abstract Background To compare the difference in central corneal thickness (CCT) measurements in normal eyes between a rotating Scheimpflug camera combined with a Placido-disk corneal topographer (Sirius, CSO, Italy) and ultrasound pachymetry (USP). Methods A systematic literature search was conducted for relevant studies published on PubMed, Medline, EMBASE, and the Cochrane Library and ClinicalTrials.gov from inception to August 1st, 2019. Primary outcome measures were CCT measurements between Sirius and USP. A random effects model was used to pool CCT measurements. Results A total of twelve studies involving 862 eyes were included in this meta-analysis. The meta-analysis found CCT measurements between Sirius and USP to be statistically significantly different (P < 0.0001). The mean difference between Sirius and USP was −11.26 μm with a 95% confidence interval (CI) (−16.92 μm, −5.60 μm). The heterogeneity was I2 = 60% (P = 0.004). Conclusion CCT measurements with the Sirius Scheimpflug-Placido topographer were statistically significantly lower than USP. However, it may be argued that the mean difference of 11.26 μm is not a clinically significant difference.
topic Meta-analysis
Central corneal thickness
Scheimpflug-Placido topographer
Ultrasound Pachymetry
url https://doi.org/10.1186/s40662-021-00227-5
work_keys_str_mv AT yilijin siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT colmmcalinden siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT yongsun siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT daizongwen siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT yiranwang siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT jinjinyu siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT kefeng siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT benhaosong siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT qinmeiwang siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT shihaochen siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
AT jinhaihuang siriusscheimpflugplacidoversusultrasoundpachymetryforcentralcornealthicknessmetaanalysis
_version_ 1724258151264419840