Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation

A patient with a dual chamber pacemaker was admitted to the emergency room after out-of-hospital cardioversion for syncopal sustained monomorphic ventricular tachycardia. Device interrogation revealed an abnormally timed ventricular spike after a ventricular premature beat at the beginning of the ev...

Full description

Bibliographic Details
Main Authors: David Roque, Nuno Cabanelas, João Augusto, Francisco Madeira, Hugo Vasconcelos, Carlos Morais
Format: Article
Language:English
Published: Elsevier 2020-10-01
Series:Revista Portuguesa de Cardiologia
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S0870255120303838
id doaj-7fa43c87b9df4e588e6d64fc3ab321ac
record_format Article
spelling doaj-7fa43c87b9df4e588e6d64fc3ab321ac2020-11-25T03:37:09ZengElsevierRevista Portuguesa de Cardiologia0870-25512020-10-013910611.e1611.e3Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillationDavid Roque0Nuno Cabanelas1João Augusto2Francisco Madeira3Hugo Vasconcelos4Carlos Morais5Corresponding author.; Serviço de Cardiologia, Hospital Professor Doutor Fernando da Fonseca, EPE, Amadora, PortugalServiço de Cardiologia, Hospital Professor Doutor Fernando da Fonseca, EPE, Amadora, PortugalServiço de Cardiologia, Hospital Professor Doutor Fernando da Fonseca, EPE, Amadora, PortugalServiço de Cardiologia, Hospital Professor Doutor Fernando da Fonseca, EPE, Amadora, PortugalServiço de Cardiologia, Hospital Professor Doutor Fernando da Fonseca, EPE, Amadora, PortugalServiço de Cardiologia, Hospital Professor Doutor Fernando da Fonseca, EPE, Amadora, PortugalA patient with a dual chamber pacemaker was admitted to the emergency room after out-of-hospital cardioversion for syncopal sustained monomorphic ventricular tachycardia. Device interrogation revealed an abnormally timed ventricular spike after a ventricular premature beat at the beginning of the event, caused by a pacemaker algorithm designed to avoid atrial fibrillation, non-competitive atrial pacing. Despite the absence of significant coronary lesions, in the setting of a vulnerable substrate – a hypokinetic and hyperechogenic region of ventricular myocardium – an upgrade to a dual-chamber implantable cardioverter-defibrillator was performed, and substrate ablation was planned. Resumo: Doente com pacemaker de dupla câmara é admitido no Serviço de Urgência após uma cardioversão fora do hospital por episódio sincopal secundário a taquicardia ventricular monomórfica mantida. A interrogação do dispositivo revelou um spike ventricular inapropriado após uma extrassístole ventricular no início do evento, provocado por um algoritmo de pacemaker desenhado para evitar fibrilhação auricular – o non-competitive atrial pacing (NCAP). Apesar da ausência de lesões coronárias significativas, no contexto de substrato vulnerável – uma região hipocinética e hiperecogénica de miocárdio ventricular – upgrade para cardiodesfibrilhador implantável (CDI) bicamaral foi realizado e ablação de substrato planeada.http://www.sciencedirect.com/science/article/pii/S0870255120303838Taquicardia ventricularAlgoritmo de pacemakerPacing auricular não competitivo
collection DOAJ
language English
format Article
sources DOAJ
author David Roque
Nuno Cabanelas
João Augusto
Francisco Madeira
Hugo Vasconcelos
Carlos Morais
spellingShingle David Roque
Nuno Cabanelas
João Augusto
Francisco Madeira
Hugo Vasconcelos
Carlos Morais
Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation
Revista Portuguesa de Cardiologia
Taquicardia ventricular
Algoritmo de pacemaker
Pacing auricular não competitivo
author_facet David Roque
Nuno Cabanelas
João Augusto
Francisco Madeira
Hugo Vasconcelos
Carlos Morais
author_sort David Roque
title Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation
title_short Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation
title_full Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation
title_fullStr Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation
title_full_unstemmed Ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation
title_sort ventricular tachycardia induced by pacing algorithm designed to avoid atrial fibrillation
publisher Elsevier
series Revista Portuguesa de Cardiologia
issn 0870-2551
publishDate 2020-10-01
description A patient with a dual chamber pacemaker was admitted to the emergency room after out-of-hospital cardioversion for syncopal sustained monomorphic ventricular tachycardia. Device interrogation revealed an abnormally timed ventricular spike after a ventricular premature beat at the beginning of the event, caused by a pacemaker algorithm designed to avoid atrial fibrillation, non-competitive atrial pacing. Despite the absence of significant coronary lesions, in the setting of a vulnerable substrate – a hypokinetic and hyperechogenic region of ventricular myocardium – an upgrade to a dual-chamber implantable cardioverter-defibrillator was performed, and substrate ablation was planned. Resumo: Doente com pacemaker de dupla câmara é admitido no Serviço de Urgência após uma cardioversão fora do hospital por episódio sincopal secundário a taquicardia ventricular monomórfica mantida. A interrogação do dispositivo revelou um spike ventricular inapropriado após uma extrassístole ventricular no início do evento, provocado por um algoritmo de pacemaker desenhado para evitar fibrilhação auricular – o non-competitive atrial pacing (NCAP). Apesar da ausência de lesões coronárias significativas, no contexto de substrato vulnerável – uma região hipocinética e hiperecogénica de miocárdio ventricular – upgrade para cardiodesfibrilhador implantável (CDI) bicamaral foi realizado e ablação de substrato planeada.
topic Taquicardia ventricular
Algoritmo de pacemaker
Pacing auricular não competitivo
url http://www.sciencedirect.com/science/article/pii/S0870255120303838
work_keys_str_mv AT davidroque ventriculartachycardiainducedbypacingalgorithmdesignedtoavoidatrialfibrillation
AT nunocabanelas ventriculartachycardiainducedbypacingalgorithmdesignedtoavoidatrialfibrillation
AT joaoaugusto ventriculartachycardiainducedbypacingalgorithmdesignedtoavoidatrialfibrillation
AT franciscomadeira ventriculartachycardiainducedbypacingalgorithmdesignedtoavoidatrialfibrillation
AT hugovasconcelos ventriculartachycardiainducedbypacingalgorithmdesignedtoavoidatrialfibrillation
AT carlosmorais ventriculartachycardiainducedbypacingalgorithmdesignedtoavoidatrialfibrillation
_version_ 1724546889413558272