DENSITOMETRY-BASED FEM SIMULATIONS OF NOVEL POROUS IMPLANTS AND CORRESPONDING STRESS DISTRIBUTION AT THE PERI-IMPLANT AREA
The presented work focuses on determining the stress distribution at the peri-implant area around dental implants. A numerical analysis simulating the conditions of chewing food has been performed on a FEM model. This model has been created using anonymized real patient CT data and a dental implant...
Main Authors: | , |
---|---|
Format: | Article |
Language: | English |
Published: |
CTU Central Library
2020-03-01
|
Series: | Acta Polytechnica CTU Proceedings |
Subjects: | |
Online Access: | https://ojs.cvut.cz/ojs/index.php/APP/article/view/6389 |
id |
doaj-84d03bd634ec4b9280622c5e0ac9bf28 |
---|---|
record_format |
Article |
spelling |
doaj-84d03bd634ec4b9280622c5e0ac9bf282020-11-25T03:10:06ZengCTU Central LibraryActa Polytechnica CTU Proceedings2336-53822020-03-01260768010.14311/APP.2020.26.00765007DENSITOMETRY-BASED FEM SIMULATIONS OF NOVEL POROUS IMPLANTS AND CORRESPONDING STRESS DISTRIBUTION AT THE PERI-IMPLANT AREALuboš Řehounek0Aleš Jíra1Czech Technical University in Prague, Faculty of Civil Engineering, Department of Mechanics, Thákurova 7, 166 29 Prague 6, Czech RepublicCzech Technical University in Prague, Faculty of Civil Engineering, Department of Mechanics, Thákurova 7, 166 29 Prague 6, Czech RepublicThe presented work focuses on determining the stress distribution at the peri-implant area around dental implants. A numerical analysis simulating the conditions of chewing food has been performed on a FEM model. This model has been created using anonymized real patient CT data and a dental implant model developed at CTU. The CT data served as a 3D geometry and also as a way to construct the global matrix of stiffness of the bone material. Bone density was used as the defining parameter in determining the values of Young’s moduli of individual finite elements by the computational program (Mechanical Finder). The implant was introduced as a user-created STL file, which was imported to the computational software and situated inside the geometry of the human mandible. The results show that, as predicted, porous implants achieve higher values of minimum principal stress in the bone as opposed to homogeneous implants (13.4 MPa vs. 7.0 MPa), thus reducing stress shielding.https://ojs.cvut.cz/ojs/index.php/APP/article/view/6389dental implant, fem, mechanical finder, porous, stress shielding |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Luboš Řehounek Aleš Jíra |
spellingShingle |
Luboš Řehounek Aleš Jíra DENSITOMETRY-BASED FEM SIMULATIONS OF NOVEL POROUS IMPLANTS AND CORRESPONDING STRESS DISTRIBUTION AT THE PERI-IMPLANT AREA Acta Polytechnica CTU Proceedings dental implant, fem, mechanical finder, porous, stress shielding |
author_facet |
Luboš Řehounek Aleš Jíra |
author_sort |
Luboš Řehounek |
title |
DENSITOMETRY-BASED FEM SIMULATIONS OF NOVEL POROUS IMPLANTS AND CORRESPONDING STRESS DISTRIBUTION AT THE PERI-IMPLANT AREA |
title_short |
DENSITOMETRY-BASED FEM SIMULATIONS OF NOVEL POROUS IMPLANTS AND CORRESPONDING STRESS DISTRIBUTION AT THE PERI-IMPLANT AREA |
title_full |
DENSITOMETRY-BASED FEM SIMULATIONS OF NOVEL POROUS IMPLANTS AND CORRESPONDING STRESS DISTRIBUTION AT THE PERI-IMPLANT AREA |
title_fullStr |
DENSITOMETRY-BASED FEM SIMULATIONS OF NOVEL POROUS IMPLANTS AND CORRESPONDING STRESS DISTRIBUTION AT THE PERI-IMPLANT AREA |
title_full_unstemmed |
DENSITOMETRY-BASED FEM SIMULATIONS OF NOVEL POROUS IMPLANTS AND CORRESPONDING STRESS DISTRIBUTION AT THE PERI-IMPLANT AREA |
title_sort |
densitometry-based fem simulations of novel porous implants and corresponding stress distribution at the peri-implant area |
publisher |
CTU Central Library |
series |
Acta Polytechnica CTU Proceedings |
issn |
2336-5382 |
publishDate |
2020-03-01 |
description |
The presented work focuses on determining the stress distribution at the peri-implant area around dental implants. A numerical analysis simulating the conditions of chewing food has been performed on a FEM model. This model has been created using anonymized real patient CT data and a dental implant model developed at CTU. The CT data served as a 3D geometry and also as a way to construct the global matrix of stiffness of the bone material. Bone density was used as the defining parameter in determining the values of Young’s moduli of individual finite elements by the computational program (Mechanical Finder). The implant was introduced as a user-created STL file, which was imported to the computational software and situated inside the geometry of the human mandible. The results show that, as predicted, porous implants achieve higher values of minimum principal stress in the bone as opposed to homogeneous implants (13.4 MPa vs. 7.0 MPa), thus reducing stress shielding. |
topic |
dental implant, fem, mechanical finder, porous, stress shielding |
url |
https://ojs.cvut.cz/ojs/index.php/APP/article/view/6389 |
work_keys_str_mv |
AT lubosrehounek densitometrybasedfemsimulationsofnovelporousimplantsandcorrespondingstressdistributionattheperiimplantarea AT alesjira densitometrybasedfemsimulationsofnovelporousimplantsandcorrespondingstressdistributionattheperiimplantarea |
_version_ |
1724660601739804672 |