Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry

AIM:To compare the difference and correlation of central corneal thickness(CCT)measured by Zeiss Visante Anterior Segment Optical Coherence Tomography(Visante OCT), Heidelberg Slit-Lamp Anterior Segment Optical Coherence Tomography(SL-OCT), OrbscanⅡz Computerized Topography(OrbscanⅡz )and A-mode ult...

Full description

Bibliographic Details
Main Authors: Duo Peng, Qin-Mei Wang, Shi-Hao Chen
Format: Article
Language:English
Published: Press of International Journal of Ophthalmology (IJO PRESS) 2013-12-01
Series:Guoji Yanke Zazhi
Subjects:
Online Access:http://www.ies.net.cn/cn_publish/2013/12/201312057.pdf
id doaj-8ecaca06329c49bfadf6578468cf3db8
record_format Article
spelling doaj-8ecaca06329c49bfadf6578468cf3db82020-11-24T22:51:35ZengPress of International Journal of Ophthalmology (IJO PRESS)Guoji Yanke Zazhi1672-51232013-12-0113122545254810.3980/j.issn.1672-5123.2013.12.57Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetryDuo PengQin-Mei WangShi-Hao ChenAIM:To compare the difference and correlation of central corneal thickness(CCT)measured by Zeiss Visante Anterior Segment Optical Coherence Tomography(Visante OCT), Heidelberg Slit-Lamp Anterior Segment Optical Coherence Tomography(SL-OCT), OrbscanⅡz Computerized Topography(OrbscanⅡz )and A-mode ultrasound pachymetry.<p>METHODS:Forty-three patients with myopia(86 eyes)were included. Sequential measurements of CCT was performed by the same operator with two different anterior segment OCT devices(Visante<sup>TM</sup>, Carl Zeiss, Meditec and Heidelberg Engineering<sup>TM</sup>), OrbscanⅡz Computerized Topography(Bausch and Lomb Inc<sup>TM</sup>)and A-mode ultrasound pachymetry. The consistency of measurements was statistically analyzed by one-way ANOVA and LSD pair-wise comparisons. The differences were analyzed by Pearson correlation coefficients.<p>RESULTS:The average CCT measured by Visante OCT, SL-OCT, OrbscanⅡz, and A-mode ultrasound pachymetry were(521.05±29.88)μm,(533.73±31.83)μm,(534.02±32.96)μm, and(540.60±31.11)μm, the 95% confidence interval were(514.57~527.54),(526.83~540.65),(533.85~547.35),(536.75~544.45). There was statistically significant difference of CCT between using of Visante OCT and SL-OCT(<i>P</i>=0.009). There was statistically significant difference of CCT using of Visante OCT and A-mode ultrasound pachymetry(<i>P</i><0.01), OrbscanⅡz(<i>P</i>=0.008). The Visante OCT CCT measurements was significantly less than CCT with SL-OCT(12.69±9.32)μm. Compare with A-mode ultrasound pachymetry and OrbscanⅡz, the Visante OCT measurements significantly underestimated central corneal thickness by a mean of(19.55±8.64)μm and(12.96±12.21)μm. There was no significant difference of CCT between using of SL-OCT and A-mode ultrasound pachymetry(<i>P</i>=0.158), OrbscanⅡz(<i>P</i>=0.954). There was no significant difference of CCT between using of OrbscanⅡz and A-mode ultrasound pachymetry(<i>P</i>=0.176). There were positive correlationship among Visante OCT, SL-OCT, OrbscanⅡz, and A-mode ultrasound pachymetry in CCT measurement(<i>P</i><0.01).<p>CONCLUSION:There are excellent correlationship among Visante OCT, SL-OCT,OrbscanⅡz, and A-mode ultrasound pachymetry in CCT measurement. The data shows that the CCT obtained by SL-OCT and OrbscanⅡz do not have significant difference as compared with that from the A-mode ultrasound pachymetry. Three optical measurement devices is more likely to be accepted by patients because of its non-contact, non-invasive, and short measurement time as compared with the A-mode ultrasound pachymetry.http://www.ies.net.cn/cn_publish/2013/12/201312057.pdfVisante OCTSL-OCTOrbscanⅡzA-mode ultrasound pachymetrycentral corneal thickness
collection DOAJ
language English
format Article
sources DOAJ
author Duo Peng
Qin-Mei Wang
Shi-Hao Chen
spellingShingle Duo Peng
Qin-Mei Wang
Shi-Hao Chen
Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry
Guoji Yanke Zazhi
Visante OCT
SL-OCT
OrbscanⅡz
A-mode ultrasound pachymetry
central corneal thickness
author_facet Duo Peng
Qin-Mei Wang
Shi-Hao Chen
author_sort Duo Peng
title Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry
title_short Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry
title_full Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry
title_fullStr Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry
title_full_unstemmed Comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry
title_sort comparison of central corneal thickness measured by three optical measurement devices with ultrasound pachymetry
publisher Press of International Journal of Ophthalmology (IJO PRESS)
series Guoji Yanke Zazhi
issn 1672-5123
publishDate 2013-12-01
description AIM:To compare the difference and correlation of central corneal thickness(CCT)measured by Zeiss Visante Anterior Segment Optical Coherence Tomography(Visante OCT), Heidelberg Slit-Lamp Anterior Segment Optical Coherence Tomography(SL-OCT), OrbscanⅡz Computerized Topography(OrbscanⅡz )and A-mode ultrasound pachymetry.<p>METHODS:Forty-three patients with myopia(86 eyes)were included. Sequential measurements of CCT was performed by the same operator with two different anterior segment OCT devices(Visante<sup>TM</sup>, Carl Zeiss, Meditec and Heidelberg Engineering<sup>TM</sup>), OrbscanⅡz Computerized Topography(Bausch and Lomb Inc<sup>TM</sup>)and A-mode ultrasound pachymetry. The consistency of measurements was statistically analyzed by one-way ANOVA and LSD pair-wise comparisons. The differences were analyzed by Pearson correlation coefficients.<p>RESULTS:The average CCT measured by Visante OCT, SL-OCT, OrbscanⅡz, and A-mode ultrasound pachymetry were(521.05±29.88)μm,(533.73±31.83)μm,(534.02±32.96)μm, and(540.60±31.11)μm, the 95% confidence interval were(514.57~527.54),(526.83~540.65),(533.85~547.35),(536.75~544.45). There was statistically significant difference of CCT between using of Visante OCT and SL-OCT(<i>P</i>=0.009). There was statistically significant difference of CCT using of Visante OCT and A-mode ultrasound pachymetry(<i>P</i><0.01), OrbscanⅡz(<i>P</i>=0.008). The Visante OCT CCT measurements was significantly less than CCT with SL-OCT(12.69±9.32)μm. Compare with A-mode ultrasound pachymetry and OrbscanⅡz, the Visante OCT measurements significantly underestimated central corneal thickness by a mean of(19.55±8.64)μm and(12.96±12.21)μm. There was no significant difference of CCT between using of SL-OCT and A-mode ultrasound pachymetry(<i>P</i>=0.158), OrbscanⅡz(<i>P</i>=0.954). There was no significant difference of CCT between using of OrbscanⅡz and A-mode ultrasound pachymetry(<i>P</i>=0.176). There were positive correlationship among Visante OCT, SL-OCT, OrbscanⅡz, and A-mode ultrasound pachymetry in CCT measurement(<i>P</i><0.01).<p>CONCLUSION:There are excellent correlationship among Visante OCT, SL-OCT,OrbscanⅡz, and A-mode ultrasound pachymetry in CCT measurement. The data shows that the CCT obtained by SL-OCT and OrbscanⅡz do not have significant difference as compared with that from the A-mode ultrasound pachymetry. Three optical measurement devices is more likely to be accepted by patients because of its non-contact, non-invasive, and short measurement time as compared with the A-mode ultrasound pachymetry.
topic Visante OCT
SL-OCT
OrbscanⅡz
A-mode ultrasound pachymetry
central corneal thickness
url http://www.ies.net.cn/cn_publish/2013/12/201312057.pdf
work_keys_str_mv AT duopeng comparisonofcentralcornealthicknessmeasuredbythreeopticalmeasurementdeviceswithultrasoundpachymetry
AT qinmeiwang comparisonofcentralcornealthicknessmeasuredbythreeopticalmeasurementdeviceswithultrasoundpachymetry
AT shihaochen comparisonofcentralcornealthicknessmeasuredbythreeopticalmeasurementdeviceswithultrasoundpachymetry
_version_ 1725669050776813568