105 Inappropriate ICD Shocks in a Patient with Dilated Cardiomyopathy and Broca’s Aphasia

With a growing number of ICD recipients, device complications are seen more frequently in the clinical setting and outpatient departments. Among the most severe are ICD infections and inappropriate therapies caused by oversensing of atrial tachycardias or lead fracture. We report on a 76-year-old fe...

Full description

Bibliographic Details
Main Authors: Christian Georgi, Michael Neuß, Viviane Möller, Martin Seifert, Christian Butter
Format: Article
Language:English
Published: Hindawi Limited 2019-01-01
Series:Case Reports in Cardiology
Online Access:http://dx.doi.org/10.1155/2019/8302591