Hyperglycemia-Induced T-Wave Oversensing as a Cause of Cardiac Resynchronization Therapy (CRT) Failure
T-wave oversensing occurs when the counter starts giving dual beeps for every cardiac cycle instead of one. This usually happens when the monitoring lead displays a tall T wave, which is also sharp. R wave sensing algorithms of the devices do not sense T wave because the slow rate of the T wave is m...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Tehran University of Medical Sciences
2012-01-01
|
Series: | Journal of Tehran University Heart Center |
Subjects: | |
Online Access: | https://jthc.tums.ac.ir/index.php/jthc/article/view/238 |
Summary: | T-wave oversensing occurs when the counter starts giving dual beeps for every cardiac cycle instead of one. This usually happens when the monitoring lead displays a tall T wave, which is also sharp. R wave sensing algorithms of the devices do not sense T wave because the slow rate of the T wave is much less than that of the R wave. But the slow rate of T waves may change with time and also because of parameters like potassium levels and hyperglycemia. We present a 67-year-old female who underwent the implantation of cardiac resynchronization therapy (cardiac resynchronization and implantable cardioverter defibrilator [CRT-D]) because of severe left ventricular systolic dysfunction and ventricular dyssynchrony experienced recurrent inappropriate implantable cardioverter-defibrillator (ICD) shocks and CRT failure. Device analysis showed that the CRT failure was in consequence of T-wave oversensing due to hyperglycemia. Elimination of the T-wave oversensing after hyperglycemia control conferred good biventricular pacing and good response to CRT during a 6-month follow-up period.
|
---|---|
ISSN: | 1735-8620 2008-2371 |