AIM: To compare the repeatability and agreement between different instruments on peripheral retina refractive error measurement. <p>METHODS: Streak retinoscopy, Shin-Nippon K5001 Auto-refractor and Aberrometor(COAS)were used to measure 28 participants' central and peripheral retinal field...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Press of International Journal of Ophthalmology (IJO PRESS)
2016-04-01
|
Series: | Guoji Yanke Zazhi |
Subjects: | |
Online Access: | http://ies.ijo.cn/cn_publish/2016/4/201604054.pdf |
Summary: | AIM: To compare the repeatability and agreement between different instruments on peripheral retina refractive error measurement. <p>METHODS: Streak retinoscopy, Shin-Nippon K5001 Auto-refractor and Aberrometor(COAS)were used to measure 28 participants' central and peripheral retinal field refractive errors. <p>RESULTS: Spherical equivalent value(<i>M</i>)were repeatable on Aberrometor and Shin-Nippon K5001 Auto-refractor(mean difference≤±0.10D),with poor repeatability on streak retinoscopy. There were good agreement between Aberrometor and Shin-Nippon K5001 Auto-refractor measurements. <p>CONCLUSION: Aberrometor and Shin-Nippon K5001 Auto-refractor both are valid tools for the central and peripheral refractive error measurement. |
---|---|