Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?

OBJECTIVES: The aim of this study is to evaluate the quality of I-124 PET images with and without prompt gamma compensation (PGC) by comparing the recovery coefficients (RC), the signal to noise ratios (SNR) and the contrast to F-18 and Ga-68. Furthermore, the influence of the PGC on the quantificat...

Full description

Bibliographic Details
Main Authors: Veronika Preylowski, Susanne Schlögl, Frédéric Schoenahl, Gerhard Jörg, Samuel Samnick, Andreas K Buck, Michael Lassmann
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2013-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC3754939?pdf=render
id doaj-b33ab7199873423e9bd66f3a18d75da4
record_format Article
spelling doaj-b33ab7199873423e9bd66f3a18d75da42020-11-25T02:22:53ZengPublic Library of Science (PLoS)PLoS ONE1932-62032013-01-0188e7172910.1371/journal.pone.0071729Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?Veronika PreylowskiSusanne SchlöglFrédéric SchoenahlGerhard JörgSamuel SamnickAndreas K BuckMichael LassmannOBJECTIVES: The aim of this study is to evaluate the quality of I-124 PET images with and without prompt gamma compensation (PGC) by comparing the recovery coefficients (RC), the signal to noise ratios (SNR) and the contrast to F-18 and Ga-68. Furthermore, the influence of the PGC on the quantification and image quality is evaluated. METHODS: For measuring the image quality the NEMA NU2-2001 PET/SPECT-Phantom was used containing 6 spheres with a diameter between 10 mm and 37 mm placed in water with different levels of background activity. Each sphere was filled with the same activity concentration measured by an independently cross-calibrated dose calibrator. The "hot" sources were acquired with a full 3D PET/CT (Biograph mCT®, Siemens Medical USA). Acquisition times were 2 min for F-18 and Ga-68, and 10 min for I-124. For reconstruction an OSEM algorithm was applied. For I-124 the images were reconstructed with and without PGC. For the calculation of the RCs the activity concentrations in each sphere were determined; in addition, the influence of the background correction was studied. RESULTS: The RCs of Ga-68 are the smallest (79%). I-124 reaches similar RCs (87% with PGC, 84% without PGC) as F-18 (84%). showing that the quantification of I-124 images is similar to F-18 and slightly better than Ga-68. With background activity the contrast of the I-124 PGC images is similar to Ga-68 and F-18 scans. There was lower background activity in the I-124 images without PGC, which probably originates from an overcorrection of the scatter contribution. Consequently, the contrast without PGC was much higher than with PGC. As a consequence PGC should be used for I-124. CONCLUSIONS: For I-124 there is only a slight influence on the quantification depending on the use of the PGC. However, there are considerable differences with respect to I-124 image quality.http://europepmc.org/articles/PMC3754939?pdf=render
collection DOAJ
language English
format Article
sources DOAJ
author Veronika Preylowski
Susanne Schlögl
Frédéric Schoenahl
Gerhard Jörg
Samuel Samnick
Andreas K Buck
Michael Lassmann
spellingShingle Veronika Preylowski
Susanne Schlögl
Frédéric Schoenahl
Gerhard Jörg
Samuel Samnick
Andreas K Buck
Michael Lassmann
Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?
PLoS ONE
author_facet Veronika Preylowski
Susanne Schlögl
Frédéric Schoenahl
Gerhard Jörg
Samuel Samnick
Andreas K Buck
Michael Lassmann
author_sort Veronika Preylowski
title Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?
title_short Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?
title_full Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?
title_fullStr Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?
title_full_unstemmed Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?
title_sort is the image quality of i-124-pet impaired by an automatic correction of prompt gammas?
publisher Public Library of Science (PLoS)
series PLoS ONE
issn 1932-6203
publishDate 2013-01-01
description OBJECTIVES: The aim of this study is to evaluate the quality of I-124 PET images with and without prompt gamma compensation (PGC) by comparing the recovery coefficients (RC), the signal to noise ratios (SNR) and the contrast to F-18 and Ga-68. Furthermore, the influence of the PGC on the quantification and image quality is evaluated. METHODS: For measuring the image quality the NEMA NU2-2001 PET/SPECT-Phantom was used containing 6 spheres with a diameter between 10 mm and 37 mm placed in water with different levels of background activity. Each sphere was filled with the same activity concentration measured by an independently cross-calibrated dose calibrator. The "hot" sources were acquired with a full 3D PET/CT (Biograph mCT®, Siemens Medical USA). Acquisition times were 2 min for F-18 and Ga-68, and 10 min for I-124. For reconstruction an OSEM algorithm was applied. For I-124 the images were reconstructed with and without PGC. For the calculation of the RCs the activity concentrations in each sphere were determined; in addition, the influence of the background correction was studied. RESULTS: The RCs of Ga-68 are the smallest (79%). I-124 reaches similar RCs (87% with PGC, 84% without PGC) as F-18 (84%). showing that the quantification of I-124 images is similar to F-18 and slightly better than Ga-68. With background activity the contrast of the I-124 PGC images is similar to Ga-68 and F-18 scans. There was lower background activity in the I-124 images without PGC, which probably originates from an overcorrection of the scatter contribution. Consequently, the contrast without PGC was much higher than with PGC. As a consequence PGC should be used for I-124. CONCLUSIONS: For I-124 there is only a slight influence on the quantification depending on the use of the PGC. However, there are considerable differences with respect to I-124 image quality.
url http://europepmc.org/articles/PMC3754939?pdf=render
work_keys_str_mv AT veronikapreylowski istheimagequalityofi124petimpairedbyanautomaticcorrectionofpromptgammas
AT susanneschlogl istheimagequalityofi124petimpairedbyanautomaticcorrectionofpromptgammas
AT fredericschoenahl istheimagequalityofi124petimpairedbyanautomaticcorrectionofpromptgammas
AT gerhardjorg istheimagequalityofi124petimpairedbyanautomaticcorrectionofpromptgammas
AT samuelsamnick istheimagequalityofi124petimpairedbyanautomaticcorrectionofpromptgammas
AT andreaskbuck istheimagequalityofi124petimpairedbyanautomaticcorrectionofpromptgammas
AT michaellassmann istheimagequalityofi124petimpairedbyanautomaticcorrectionofpromptgammas
_version_ 1724861230415347712