Concurrent Validity and Reliability of an Inertial Measurement Unit for the Assessment of Craniocervical Range of Motion in Subjects with Cerebral Palsy

Objective: This study aimed to determine the validity and reliability of Inertial Measurement Units (IMUs) for the assessment of craniocervical range of motion (ROM) in patients with cerebral palsy (CP). Methods: twenty-three subjects with CP and 23 controls, aged between 4 and 14 years, were evalua...

Full description

Bibliographic Details
Main Authors: Cristina Carmona-Pérez, Juan Luis Garrido-Castro, Francisco Torres Vidal, Sandra Alcaraz-Clariana, Lourdes García-Luque, Francisco Alburquerque-Sendín, Daiana Priscila Rodrigues-de-Souza
Format: Article
Language:English
Published: MDPI AG 2020-02-01
Series:Diagnostics
Subjects:
Online Access:https://www.mdpi.com/2075-4418/10/2/80
Description
Summary:Objective: This study aimed to determine the validity and reliability of Inertial Measurement Units (IMUs) for the assessment of craniocervical range of motion (ROM) in patients with cerebral palsy (CP). Methods: twenty-three subjects with CP and 23 controls, aged between 4 and 14 years, were evaluated on two occasions, separated by 3 to 5 days. An IMU and a Cervical Range of Motion device (CROM) were used to assess craniocervical ROM in the three spatial planes. Validity was assessed by comparing IMU and CROM data using the Pearson correlation coefficient, the paired t-test and Bland&#8722;Altman plots. Intra-day and inter-day relative reliability were determined using the Intraclass Correlation Coefficient (ICC). The Standard Error of Measurement (SEM) and the Minimum Detectable Change at a 90% confidence level (MDC<sub>90</sub>) were obtained for absolute reliability. Results: High correlations were detected between methods in both groups on the sagittal and frontal planes (r &gt; 0.9), although this was reduced in the case of the transverse plane. Bland&#8722;Altman plots indicated bias below 5&#186;, although for the range of cervical rotation in the CP group, this was 8.2&#186;. The distance between the limits of agreement was over 23.5&#186; in both groups, except for the range of flexion-extension in the control group. ICCs were higher than 0.8 for both comparisons and groups, except for inter-day comparisons of rotational range in the CP group. Absolute reliability showed high variability, with most SEM below 8.5&#186;, although with worse inter-day results, mainly in CP subjects, with the MDC<sub>90</sub> of rotational range achieving more than 20&#186;. Conclusions: IMU application is highly correlated with CROM for the assessment of craniocervical movement in CP and healthy subjects; however, both methods are not interchangeable. The IMU error of measurement can be considered clinically acceptable; however, caution should be taken when this is used as a reference measure for interventions.
ISSN:2075-4418