Monte Carlo Evaluation of Gamma Knife Dose Profile in Real Brain Phantom

Introduction The Gamma Knife system is designed solely for non-invasive treatment of brain disorders, and it benefits from stereotactic surgical techniques. Dose calculations required in the system are performed by GammaPlan code; in this code, brain tissue is considered uniform. In the present stud...

Full description

Bibliographic Details
Main Authors: Ali Aghaebrahimian, Mohammad Hassan Alamatsaz
Format: Article
Language:English
Published: Mashhad University of Medical Sciences 2016-03-01
Series:Iranian Journal of Medical Physics
Subjects:
Online Access:http://ijmp.mums.ac.ir/article_7139_759f1c9ec8dc29878ed714e283b8142a.pdf
Description
Summary:Introduction The Gamma Knife system is designed solely for non-invasive treatment of brain disorders, and it benefits from stereotactic surgical techniques. Dose calculations required in the system are performed by GammaPlan code; in this code, brain tissue is considered uniform. In the present study, we evaluated the effect of Gamma Knife system on the obtained dose through simulating a real human brain phantom. Materials and Methods In this study, a Monte Carlo simulation code (MCNPX2.7) was employed to simulate Gamma Knife system. Brain tissue equivalent Snyder phantom and combinations were considered according to International Commission on Radiological Units (ICRU)-44 report. Results To ensure accuracy of the simulations, patient’s head was modeled by a spherical water phantom. At this point, the dosimetry parameters were compared with those obtained by the Monte Carlo code EGS4 and good consistency was observed (less than 7% difference). At the next stage, the above dosimetry parameters were compared with those obtained experimentally by polystyrene phantom and EDR2 dosimetry film and improved consistency was detected (less than 0.5% difference). Finally, the Snyder phantom, as the human brain, was simulated. The Full Width at Half Maximum (FWHM) and penumbra decreased by 4.7% and 18%, respectively. Moreover, an isocenter dose reduction of 30-40%, compared to the water phantom, was noted. Conclusion The calculation of the real phantom showed that water and polystyrene could function similarly, while evaluating dosimetry parameters in the Gamma Knife system; thus, water and polystyrene are not appropriate phantom matters for this purpose.
ISSN:2345-3672
2345-3672