Calibration of digital wide-range neutron power measurement channel for open-pool type research reactor

As the modernization of the nuclear instrumentation system progresses, research reactors have adopted digital wide-range neutron power measurement (DWRNPM) systems. These systems typically monitor the neutron flux across a range of over 10 decades. Because neutron detectors only measure the local ne...

Full description

Bibliographic Details
Main Authors: Sungmoon Joo, Jong Bok Lee, Sang Mun Seo
Format: Article
Language:English
Published: Elsevier 2018-02-01
Series:Nuclear Engineering and Technology
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S173857331730325X
id doaj-be37e7a26f6a462c818069ab18384d04
record_format Article
spelling doaj-be37e7a26f6a462c818069ab18384d042020-11-24T23:21:57ZengElsevierNuclear Engineering and Technology1738-57332018-02-0150120321010.1016/j.net.2017.10.011Calibration of digital wide-range neutron power measurement channel for open-pool type research reactorSungmoon Joo0Jong Bok Lee1Sang Mun Seo2Decommissioning Technology Research Division, Korea Atomic Energy Research Institute, 111 Daedeok-daero 989Beon-gil, Daejeon 34075, Republic of KoreaResearch Reactor System Design Division, Korea Atomic Energy Research Institute, 111 Daedeok-daero 989Beon-gil, Daejeon 34075, Republic of KoreaResearch Reactor System Design Division, Korea Atomic Energy Research Institute, 111 Daedeok-daero 989Beon-gil, Daejeon 34075, Republic of KoreaAs the modernization of the nuclear instrumentation system progresses, research reactors have adopted digital wide-range neutron power measurement (DWRNPM) systems. These systems typically monitor the neutron flux across a range of over 10 decades. Because neutron detectors only measure the local neutron flux at their position, the local neutron flux must be converted to total reactor power through calibration, which involves mapping the local neutron flux level to a reference reactor power. Conventionally, the neutron power range is divided into smaller subranges because the neutron detector signal characteristics and the reference reactor power estimation methods are different for each subrange. Therefore, many factors should be considered when preparing the calibration procedure for DWRNPM channels. The main purpose of this work is to serve as a reference for performing the calibration of DWRNPM systems in research reactors. This work provides a comprehensive overview of the calibration of DWRNPM channels by describing the configuration of the DWRNPM system and by summarizing the theories of operation and the reference power estimation methods with their associated calibration procedure. The calibration procedure was actually performed during the commissioning of an open-pool type research reactor, and the results and experience are documented herein.http://www.sciencedirect.com/science/article/pii/S173857331730325XCommissioningDigital Wide-Range Neutron Power MeasurementFission ChamberNuclear Instrumentation SystemPower CalibrationResearch Reactor
collection DOAJ
language English
format Article
sources DOAJ
author Sungmoon Joo
Jong Bok Lee
Sang Mun Seo
spellingShingle Sungmoon Joo
Jong Bok Lee
Sang Mun Seo
Calibration of digital wide-range neutron power measurement channel for open-pool type research reactor
Nuclear Engineering and Technology
Commissioning
Digital Wide-Range Neutron Power Measurement
Fission Chamber
Nuclear Instrumentation System
Power Calibration
Research Reactor
author_facet Sungmoon Joo
Jong Bok Lee
Sang Mun Seo
author_sort Sungmoon Joo
title Calibration of digital wide-range neutron power measurement channel for open-pool type research reactor
title_short Calibration of digital wide-range neutron power measurement channel for open-pool type research reactor
title_full Calibration of digital wide-range neutron power measurement channel for open-pool type research reactor
title_fullStr Calibration of digital wide-range neutron power measurement channel for open-pool type research reactor
title_full_unstemmed Calibration of digital wide-range neutron power measurement channel for open-pool type research reactor
title_sort calibration of digital wide-range neutron power measurement channel for open-pool type research reactor
publisher Elsevier
series Nuclear Engineering and Technology
issn 1738-5733
publishDate 2018-02-01
description As the modernization of the nuclear instrumentation system progresses, research reactors have adopted digital wide-range neutron power measurement (DWRNPM) systems. These systems typically monitor the neutron flux across a range of over 10 decades. Because neutron detectors only measure the local neutron flux at their position, the local neutron flux must be converted to total reactor power through calibration, which involves mapping the local neutron flux level to a reference reactor power. Conventionally, the neutron power range is divided into smaller subranges because the neutron detector signal characteristics and the reference reactor power estimation methods are different for each subrange. Therefore, many factors should be considered when preparing the calibration procedure for DWRNPM channels. The main purpose of this work is to serve as a reference for performing the calibration of DWRNPM systems in research reactors. This work provides a comprehensive overview of the calibration of DWRNPM channels by describing the configuration of the DWRNPM system and by summarizing the theories of operation and the reference power estimation methods with their associated calibration procedure. The calibration procedure was actually performed during the commissioning of an open-pool type research reactor, and the results and experience are documented herein.
topic Commissioning
Digital Wide-Range Neutron Power Measurement
Fission Chamber
Nuclear Instrumentation System
Power Calibration
Research Reactor
url http://www.sciencedirect.com/science/article/pii/S173857331730325X
work_keys_str_mv AT sungmoonjoo calibrationofdigitalwiderangeneutronpowermeasurementchannelforopenpooltyperesearchreactor
AT jongboklee calibrationofdigitalwiderangeneutronpowermeasurementchannelforopenpooltyperesearchreactor
AT sangmunseo calibrationofdigitalwiderangeneutronpowermeasurementchannelforopenpooltyperesearchreactor
_version_ 1725569312983351296