Comparison of Wave Amplitude, Implicit Time and Comfort Level Using Dawson-Trick-Litzkow, Jet and Dencott Electrodes in Electroretinography in Normal Adults

Abstract   Introduction Several electrodes can be used in standard full-field electroretinography (ERG). However, there are no standard values for Indonesians yet. Therefore, this research aims to establish normal values, to compare them and to compare comfort levels using DTL, Dencott and...

Full description

Bibliographic Details
Main Authors: Theresia Yinski Pistari Gondosari, Mohamad Sidik, Syntia Nusanti
Format: Article
Language:English
Published: International Journal of Retina 2020-09-01
Series:IJRETINA (International Journal of Retina)
Online Access:https://www.ijretina.com/index.php/ijretina/article/view/123
Description
Summary:Abstract   Introduction Several electrodes can be used in standard full-field electroretinography (ERG). However, there are no standard values for Indonesians yet. Therefore, this research aims to establish normal values, to compare them and to compare comfort levels using DTL, Dencott and Jet electrodes in Indonesian adults.   Method Through convenient sampling, 58 normal Indonesian subjects, age 19-49 years old were selected. ERG amplitudes and implicit time values were measured according to recommendations by the International Society for Clinical Electrophysiology of Vision (ISCEV). Evaluations consisted of scotopic 0.01, 3.0, OP and photopic 3.0 flicker. After examination, all subjects filled in a questionnaire about comfort levels, adopted from the visual analog scale.   Result We observed a statistically significant difference in ERG normal values between electrodes with higher wave amplitudes and longer implicit times in Dencott and jet electrodes, compared to DTL electrodes. Jet and DTL electrodes are more comfortable than Dencott electrodes for Indonesian adults in standard full-field electroretinography.   Conclusion DTL electrodes give the lowest wave amplitudes and the shortest implicit times and are the most comfortable electrode compared to Dencott and ERG jet electrodes, in standard full-field electroretinography in Indonesian adults.   Keywords: electroretinography, Dencott, DTL, Jet, electrode    
ISSN:2614-8684
2614-8536