Design of a preamplifier card for the photomultiplier tubes of a gamma camera
The service provided by Gamma Cameras (GC) in Nuclear Medicine departments fails because of their breakdown, generally due to the associated electronics and not to the physical detection components. Therefore, it was decided to develop an electronic system that allows the recovery and optimization o...
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Centro de Gestión de la Información y Desarrollo de la Energía (CUBAENERGIA)
2019-08-01
|
Series: | Nucleus |
Subjects: | |
Online Access: | http://nucleus.cubaenergia.cu/index.php/nucleus/article/view/677 |
id |
doaj-cfe6560d7f8d4272ad1406a85be8de99 |
---|---|
record_format |
Article |
spelling |
doaj-cfe6560d7f8d4272ad1406a85be8de992020-11-25T03:06:12ZengCentro de Gestión de la Información y Desarrollo de la Energía (CUBAENERGIA)Nucleus0864-084X2075-56352019-08-010653641675Design of a preamplifier card for the photomultiplier tubes of a gamma cameraJorge Luis Domínguez MartínezDoris del Carmen Rivero RamírezJuan Francisco Osorio DelizThe service provided by Gamma Cameras (GC) in Nuclear Medicine departments fails because of their breakdown, generally due to the associated electronics and not to the physical detection components. Therefore, it was decided to develop an electronic system that allows the recovery and optimization of disused GC, starting with the design of the preamplifier for each photomultiplier tube (PMT). The circuit was designed and simulated and the list of components necessary for the construction of the preamplifier was generated, as well as the printed circuit board was designed for its assembly. By simulating the preamplifier, this worked in linear mode. This determines that the amplitude of the output signal is proportional to the amount of charge delivered by the detector. This card allows an automatic adjustment of the signals of the PMTs as modern GC do. Besides, the circuit was designed and simulated for 37 and 75 PMTs, and the printed circuit board was designed for both cases.http://nucleus.cubaenergia.cu/index.php/nucleus/article/view/677medicina nuclearcubacámaras gammafotomultiplicadoresamplificadores |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Jorge Luis Domínguez Martínez Doris del Carmen Rivero Ramírez Juan Francisco Osorio Deliz |
spellingShingle |
Jorge Luis Domínguez Martínez Doris del Carmen Rivero Ramírez Juan Francisco Osorio Deliz Design of a preamplifier card for the photomultiplier tubes of a gamma camera Nucleus medicina nuclear cuba cámaras gamma fotomultiplicadores amplificadores |
author_facet |
Jorge Luis Domínguez Martínez Doris del Carmen Rivero Ramírez Juan Francisco Osorio Deliz |
author_sort |
Jorge Luis Domínguez Martínez |
title |
Design of a preamplifier card for the photomultiplier tubes of a gamma camera |
title_short |
Design of a preamplifier card for the photomultiplier tubes of a gamma camera |
title_full |
Design of a preamplifier card for the photomultiplier tubes of a gamma camera |
title_fullStr |
Design of a preamplifier card for the photomultiplier tubes of a gamma camera |
title_full_unstemmed |
Design of a preamplifier card for the photomultiplier tubes of a gamma camera |
title_sort |
design of a preamplifier card for the photomultiplier tubes of a gamma camera |
publisher |
Centro de Gestión de la Información y Desarrollo de la Energía (CUBAENERGIA) |
series |
Nucleus |
issn |
0864-084X 2075-5635 |
publishDate |
2019-08-01 |
description |
The service provided by Gamma Cameras (GC) in Nuclear Medicine departments fails because of their breakdown, generally due to the associated electronics and not to the physical detection components. Therefore, it was decided to develop an electronic system that allows the recovery and optimization of disused GC, starting with the design of the preamplifier for each photomultiplier tube (PMT). The circuit was designed and simulated and the list of components necessary for the construction of the preamplifier was generated, as well as the printed circuit board was designed for its assembly. By simulating the preamplifier, this worked in linear mode. This determines that the amplitude of the output signal is proportional to the amount of charge delivered by the detector. This card allows an automatic adjustment of the signals of the PMTs as modern GC do. Besides, the circuit was designed and simulated for 37 and 75 PMTs, and the printed circuit board was designed for both cases. |