Image guidance and positioning accuracy in clinical practice: influence of positioning errors and imaging dose on the real dose distribution for head and neck cancer treatment

Abstract Background Modern radiotherapy offers the possibility of highly accurate tumor treatment. To benefit from this precision at its best, regular positioning verification is necessary. By the use of image-guided radiotherapy and the application of safety margins the influence of positioning ina...

Full description

Bibliographic Details
Main Authors: Katharina Bell, Norbert Licht, Christian Rübe, Yvonne Dzierma
Format: Article
Language:English
Published: BMC 2018-10-01
Series:Radiation Oncology
Subjects:
Online Access:http://link.springer.com/article/10.1186/s13014-018-1141-8
Description
Summary:Abstract Background Modern radiotherapy offers the possibility of highly accurate tumor treatment. To benefit from this precision at its best, regular positioning verification is necessary. By the use of image-guided radiotherapy and the application of safety margins the influence of positioning inaccuracies can be counteracted. In this study the effect of additional imaging dose by set-up verification is compared with the effect of dose smearing by positioning inaccuracies for a collective of head-and-neck cancer patients. Methods This study is based on treatment plans of 40 head-and-neck cancer patients. To evaluate the imaging dose several image guidance scenarios with different energies, techniques and frequencies were simulated and added to the original plan. The influence of the positioning inaccuracies was assessed by the use of real applied table shifts for positioning. The isocenters were shifted back appropriately to these values to simulate that no positioning correction had been performed. For the single fractions the shifted plans were summed considering three different scenarios: The summation of only shifted plans, the consideration of the original plan for the fractions with set-up verification, and the addition of the extra imaging dose to the latter. For both effects (additional imaging dose and dose smearing), plans were analyzed and compared considering target coverage, sparing of organs at risk (OAR) and normal tissue complication probability (NTCP). Results Daily verification of the patient positioning using 3D imaging with MV energies result in non-negligible high doses. kV imaging has only marginal influence on plan quality, primarily related to sparing of organs at risk, even with daily 3D imaging. For this collective, sparing of organs at risk and NTCP are worse due to potential positioning errors. Conclusion Regular set-up verification is essential for precise radiation treatment. Relating to the additional dose, the use of kV modalities is uncritical for any frequency and technique. Dose smearing due to positioning errors for this collective mainly resulted in a decrease of OAR sparing. Target coverage also suffered from the positioning inaccuracies, especially for individual patients. Taking into account both examined effects the relevance of an extensive IGRT is clearly present, even at the expense of additional imaging dose and time expenditure.
ISSN:1748-717X