High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
Abstract A 46‐year‐old female with a body mass index of 38.9 kg/m2 and no organic heart disease underwent a subcutaneous implantable cardioverter‐defibrillator implantation for secondary prevention of sudden cardiac death in the setting of idiopathic ventricular fibrillation. Defibrillation threshol...