Validation of Method Performance of pH, PCO2 , PO2 , Na+, K+ of Cobas b121 ABG Analyser

Introduction: The introduction of a new method or new analyser is a common occurrence in clinical biochemistry laboratory. Blood gas measurements and electrolytes are often performed in Point-of-Care (POC) settings. When a new POC analyser is obtained, the performance of the analyser should be e...

Full description

Bibliographic Details
Main Authors: Sunil Kumar Nanda, Lopamudra Ray, Asha Dinakaran
Format: Article
Language:English
Published: JCDR Research and Publications Private Limited 2014-06-01
Series:Journal of Clinical and Diagnostic Research
Subjects:
Online Access:https://jcdr.net/articles/PDF/4442/9218_CE(Ra1)_F(H)_PF1(AGAK)_PFA(AK)_PF2(AGAK)_PF2(PN).pdf
id doaj-edc6765f90374ec8a6871fcf339cca17
record_format Article
spelling doaj-edc6765f90374ec8a6871fcf339cca172020-11-25T03:13:21ZengJCDR Research and Publications Private LimitedJournal of Clinical and Diagnostic Research2249-782X0973-709X2014-06-0186CC05CC0710.7860/JCDR/2014/9218.4442Validation of Method Performance of pH, PCO2 , PO2 , Na+, K+ of Cobas b121 ABG AnalyserSunil Kumar Nanda0Lopamudra Ray1Asha Dinakaran2Associate Professor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Affiliated to Pondicherry University, Ganapathichettikulam, Kalapet, Pondicherry, India.Assistant Professor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Affiliated to Pondicherry University, Ganapathichettikulam, Kalapet, Pondicherry, India.Tutor, Department of Biochemistry, Pondicherry Institute of Medical Sciences, Affiliated to Pondicherry University, Ganapathichettikulam, Kalapet, Pondicherry, India.Introduction: The introduction of a new method or new analyser is a common occurrence in clinical biochemistry laboratory. Blood gas measurements and electrolytes are often performed in Point-of-Care (POC) settings. When a new POC analyser is obtained, the performance of the analyser should be evaluated by comparison to the measurements with the reference analyser in the laboratory. Objectives: Evaluation of method performance of pH, PCO2 , PO2 , Na+, K+ of cobas b121 ABG analyser. Materials & Methods: The evaluation of method performance of pH, PO2 , PCO2 , Na+, K+ of cobas b121 ABG analyser was done by comparing the results of 50 patient samples run on cobas b121 with the results obtained from Rapid lab values (reference analyser). Correlation coefficient was calculated from the results obtained from both the analysers. Precision was calculated by running biorad ABG control samples. Results: The correlation coefficient values obtained for parameters were close to 1.0 indicating good correlation. The CV obtained for all the parameters were less than 5 indicating good precision. Conclusion: The new ABG analyser, Cobas b121 correlated well with the reference ABG analyser (Rapid Lab) and could be used to run on patient samples.https://jcdr.net/articles/PDF/4442/9218_CE(Ra1)_F(H)_PF1(AGAK)_PFA(AK)_PF2(AGAK)_PF2(PN).pdfblood gaselectrolytespoint of care
collection DOAJ
language English
format Article
sources DOAJ
author Sunil Kumar Nanda
Lopamudra Ray
Asha Dinakaran
spellingShingle Sunil Kumar Nanda
Lopamudra Ray
Asha Dinakaran
Validation of Method Performance of pH, PCO2 , PO2 , Na+, K+ of Cobas b121 ABG Analyser
Journal of Clinical and Diagnostic Research
blood gas
electrolytes
point of care
author_facet Sunil Kumar Nanda
Lopamudra Ray
Asha Dinakaran
author_sort Sunil Kumar Nanda
title Validation of Method Performance of pH, PCO2 , PO2 , Na+, K+ of Cobas b121 ABG Analyser
title_short Validation of Method Performance of pH, PCO2 , PO2 , Na+, K+ of Cobas b121 ABG Analyser
title_full Validation of Method Performance of pH, PCO2 , PO2 , Na+, K+ of Cobas b121 ABG Analyser
title_fullStr Validation of Method Performance of pH, PCO2 , PO2 , Na+, K+ of Cobas b121 ABG Analyser
title_full_unstemmed Validation of Method Performance of pH, PCO2 , PO2 , Na+, K+ of Cobas b121 ABG Analyser
title_sort validation of method performance of ph, pco2 , po2 , na+, k+ of cobas b121 abg analyser
publisher JCDR Research and Publications Private Limited
series Journal of Clinical and Diagnostic Research
issn 2249-782X
0973-709X
publishDate 2014-06-01
description Introduction: The introduction of a new method or new analyser is a common occurrence in clinical biochemistry laboratory. Blood gas measurements and electrolytes are often performed in Point-of-Care (POC) settings. When a new POC analyser is obtained, the performance of the analyser should be evaluated by comparison to the measurements with the reference analyser in the laboratory. Objectives: Evaluation of method performance of pH, PCO2 , PO2 , Na+, K+ of cobas b121 ABG analyser. Materials & Methods: The evaluation of method performance of pH, PO2 , PCO2 , Na+, K+ of cobas b121 ABG analyser was done by comparing the results of 50 patient samples run on cobas b121 with the results obtained from Rapid lab values (reference analyser). Correlation coefficient was calculated from the results obtained from both the analysers. Precision was calculated by running biorad ABG control samples. Results: The correlation coefficient values obtained for parameters were close to 1.0 indicating good correlation. The CV obtained for all the parameters were less than 5 indicating good precision. Conclusion: The new ABG analyser, Cobas b121 correlated well with the reference ABG analyser (Rapid Lab) and could be used to run on patient samples.
topic blood gas
electrolytes
point of care
url https://jcdr.net/articles/PDF/4442/9218_CE(Ra1)_F(H)_PF1(AGAK)_PFA(AK)_PF2(AGAK)_PF2(PN).pdf
work_keys_str_mv AT sunilkumarnanda validationofmethodperformanceofphpco2po2nakofcobasb121abganalyser
AT lopamudraray validationofmethodperformanceofphpco2po2nakofcobasb121abganalyser
AT ashadinakaran validationofmethodperformanceofphpco2po2nakofcobasb121abganalyser
_version_ 1724647318782738432