Toward Clinically Compatible Phase-Contrast Mammography.

Phase-contrast mammography using laboratory X-ray sources is a promising approach to overcome the relatively low sensitivity and specificity of clinical, absorption-based screening. Current research is mostly centered on identifying potential diagnostic benefits arising from phase-contrast and dark-...

Full description

Bibliographic Details
Main Authors: Kai Scherer, Konstantin Willer, Lukas Gromann, Lorenz Birnbacher, Eva Braig, Susanne Grandl, Anikó Sztrókay-Gaul, Julia Herzen, Doris Mayr, Karin Hellerhoff, Franz Pfeiffer
Format: Article
Language:English
Published: Public Library of Science (PLoS) 2015-01-01
Series:PLoS ONE
Online Access:http://europepmc.org/articles/PMC4481352?pdf=render
Description
Summary:Phase-contrast mammography using laboratory X-ray sources is a promising approach to overcome the relatively low sensitivity and specificity of clinical, absorption-based screening. Current research is mostly centered on identifying potential diagnostic benefits arising from phase-contrast and dark-field mammography and benchmarking the latter with conventional state-of-the-art imaging methods. So far, little effort has been made to adjust this novel imaging technique to clinical needs. In this article, we address the key points for a successful implementation to a clinical routine in the near future and present the very first dose-compatible and rapid scan-time phase-contrast mammograms of both a freshly dissected, cancer-bearing mastectomy specimen and a mammographic accreditation phantom.
ISSN:1932-6203